11 Maintenance
❚
Obey the maintenance intervals claimed in chapter “maintenance”.
❚
Obey the competences claimed in chapter “maintenance”.
❚
A second person has to be present during maintenance.
❚
reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing.
❚
❚
personnel via telephone and writing.
Do not perform maintenance during use.
Proper maintenance is an important pre-condition for safety, reliability, function and accuracy of the device.
Support h/p/cosmos service personnel with the documents needed.
11.1 Intervals and competences
Maintenance
Interval
Competence
Daily inspection
Daily
Operator
Lubrication
When OIL message occurs
Operator
Adjustment of running belt
If due
Operator
Safety inspection
12 months
h/p/cosmos service personnel only
Change of safety arch rope
24 months
h/p/cosmos service personnel only
Tightening of the running belt
If due
h/p/cosmos service personnel only
Adjustment of levelling sockets
If due
h/p/cosmos service personnel only
Installation and repair work
If due
h/p/cosmos service personnel only
h/p/cosmos recommends entering into a service contract with an authorized h/p/cosmos service technician.
A service contract provides the best preventive maintenance and care for the device.
The service contract is available at
.
To receive information on becoming h/p/cosmos service personnel, please contact [email protected].
11.2 Daily inspection
❚
Perform a daily visual inspection (see chapter “maintenance”).
Before daily use, check the whole device for wear and tear.
Description
Illustration
Pay special attention to components with high probability of wear
and tear:
1. Running belt and non-slip surfaces
2. External cabling
3. All textile parts
4. Fall prevention incl. safety harness, buckle, carabiner, rope
and rope arrest
If there is any visible or assumed wear and tear, unplug device,
exclude reconnection and call h/p/cosmos service personnel.