background image

8

          

               

        Figure 6 Work Tip                    Figure 7 Installation of Work Tip

         

                          

open

      Figure 8 Installation of 

           Work Tip Protector                            Figure 9 Error code

4.3 Installation and replacement of battery

When replacing the battery, as shown in Figure 10, first rotate the 

battery barrel counterclockwise to remove the battery tube, then take the 

old battery out of the battery tube, replace it with a new one, and finally 

tighten the battery tube clockwise according to the corresponding thread.

 Warning: 

When removing the battery, the screw under the 

battery barrel (pointed by the arrow in Figure 10) does not need to be 

unscrewed,just push the connector slightly inward to remove the battery. 

Improper replacement of lithium batteries may result in unacceptable 

risks,so replacement of lithium batteries requires trained personnel.

         

        

Figure 10 Replacement of battery

5 Operation method

1. According to the situation of patient, select suitable Work Tip and 

install it. When installing the Work Tip, chose a suitable angle to install 

the Work Tip.

 Danger:

Don’t turn on the device when installing the Work Tip, to prevent 

scalding the user by mistakenly pressing the heating button.

2. After pressing the “ON/OFF” button, the display screen of Heating 

and Packing Instrument lights up and display the preheating temperature 

and power status.

Содержание 28-780

Страница 1: ......

Страница 2: ...lation and disassembly method of accessories 7 5 Operation method 8 6 Charging instruction 9 7 Safety precautions 9 8 Cleaning Disinfection Sterilization and Maintenance 10 9 Troubleshooting 16 10 After sales service 17 11 Environment protection 17 12 EMC Declaration of comformity 18 13 Statement 22 ...

Страница 3: ...eration please read this Instruction Manual carefully before use Depending on the level of risk involved safety requirements are classed under the following indications Danger always referred to personal injury Warning referred to possible damage to property 1 Product introduction 1 1 Intended use It is used to provide heat to the Work Tip cut the gutta percha point and soften and pressurize the g...

Страница 4: ...own Under ON state shortly press ON OFF button to change the preset temperature of Work Tip The preset temperature will change to the next with the sequence 150 180 200 230 after each press And then go back to 150 after short press at temperature of 230 150 180 200 230 Figure 1 Preset temperature 2 Heating button Under the ON state connect the Work Tip and press Heating button to start heating Rel...

Страница 5: ...isplay screen shows the preset temperature value About 1s after the temperature preset the OLED screen will display the real time temperature of the Work Tip When the Heating and Packing Instrument is in the heating state the temperature indicator will simultaneously display the current temperature of the Work Tip 6 Charging base Firstly connect the power adapter plug to the power adapter as shown...

Страница 6: ...battery level of the Heating and Packing Instrument led display screen is the highest After the battery runs out the time of battery charging takes at least 2 hours and 30 minutes Figure 3 Installation of power adapter Figure 4 Connection to power supply Figure 5 Charging 1 3 Device includes 1 Heating and Packing Instrument 2 Charging base 3 Power adapter with cord 4 Work Tips The models are as sh...

Страница 7: ...50 180 200 230 e If there is no operation for 10 minutes the Heating and Packing Instrument will automatically shut down 1 4 2 Scope of application Used in the root canal obturation stage in endodontic treatment 1 5 Product specifications Size Heating and Packing Instrument 23 8mm 158 3mm 23 8mm Charging base 75 5mm 149 7mm 62 6mm Weight Heating and Packing Instrument 80g Charging base 195g Power ...

Страница 8: ...midity is 10 93 the air pressure is 70kPa 106kP and the temperature is 20 55 4 Please avoid the device from strong shock or vibration during transport And please handle it carefully 5 Please do not mix the device with hazardous articles during transport 6 Please avoid the device from sun rain and snow during transport 2 Standard icons Product serial number Follow Instructions for Use Manufacturer ...

Страница 9: ...standard of this power adapter Please install in accordance with the procedures in Figure 3 Figure 4 and Figure 5 Note The installation in Figure 3 had been finished before delivery 4 2 Installation and removal of Work Tip 1 After turning off the power switch you can directly pull the Work Tip off the Heating and Packing Instrument 2 Place the used Work Tip in a certain container and disinfect it ...

Страница 10: ...row in Figure 10 does not need to be unscrewed just push the connector slightly inward to remove the battery Improper replacement of lithium batteries may result in unacceptable risks so replacement of lithium batteries requires trained personnel Figure 10 Replacement of battery 5 Operation method 1 According to the situation of patient select suitable Work Tip and install it When installing the W...

Страница 11: ... in this product has no memory and can be used at any time or charged at any time 6 3 Before first use of this device please charge it at least for 3 hours Warning Only unplug the adapter to disconnect from the network power 7 Safety precautions 1 Do not polish the Work Tip 2 Do not knock or scratch the Heating and Packing Instrument 3 Do not place the Heating and Packing Instrument near an electr...

Страница 12: ... Work Tip Protector are as follow Unless otherwise stated they will be hereinafter referred to as products Warnings The use of strong detergent and disinfectant alkaline pH 9 or acid pH 5 will reduce the life span of products And in such cases the manufacturer takes no responsibility This device shall not be exposed to high temperature above 138 Processing limit The products have been designed for...

Страница 13: ...otes The water used here must be pure water distilled water or deionized water 8 2 Preparation before cleaning Steps Tools tray clean and dry soft cloth 1 Remove the Work Tip Protector from the handle and put it into a clean tray 2 Remove the Work Tip from the handle and place it in a clean tray 3 Wipe the Work Tip and Work Tip Protector with a soft cloth until no dirt can be seen on the surface T...

Страница 14: ...eed 20 minutes The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883 Cleaning and disinfecting steps by using Washer disinfector 1 Carefully place the product into the disinfection basket Fixation of product is needed only when the product is removable in the device The products are not allowed to contact each other 2 Start the program 3 After the program is finished ...

Страница 15: ...eps For example pure water that is in accordance with the European Pharmacopoeia or the United States Pharmacopoeia f After cleaning the chemical residue should be less than 10mg L g The air used for drying must be filtered by HEPA h Regularly repair and inspect the disinfector 8 5 Drying If your cleaning and disinfection process does not have an automatic drying function dry it after cleaning and...

Страница 16: ...SO 11607 b It can withstand high temperature of 138 C and has sufficient steam permeability c The packaging environment and related tools must be cleaned regularly to ensure cleanliness and prevent the introduction of contaminants d Avoid contact with parts of different metals when packaging 8 8 Sterilization Use only the following steam sterilization procedures fractional pre vacuum procedure for...

Страница 17: ...lated non corrosive atmosphere with a relative humidity of 10 to 93 an atmospheric pressure of 70KPa to 106KPa and a temperature of 20 C to 55 C 8 9 2 After sterilization the product should be packaged in a medical sterilization bag or a clean sealing container and stored in a special storage cabinet The storage time should not exceed 7 days If it is exceeded it should be reprocessed before use No...

Страница 18: ... self repairing spare parts If there is any abnormality in the equipment please contact our company for maintenance and do not disassemble without authorization With our company s consent we will provide circuit diagrams component part lists descriptions calibration instructions to assist to SERVICE PERSONNEL in parts repair 9 Troubleshooting Fault Cause Solution No indications no response 1 Inade...

Страница 19: ...aller Contact local distributor or manufacturer to buy new batteries for replacement OPEN code appears on display screen 1 The Work Tip is damaged 2 The Work Tip is not installed 3 The Work Tip is not well installed 1 Replace the Work Tip 2 Install the Work Tip 3 Unplug the Work Tip and reconnect If the problem still cannot be solved please contact your local dealer or our company 10 After sales s...

Страница 20: ...del Fi P should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The model Fi P uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR11 Class B The model Fi P is suitable for used in ...

Страница 21: ...and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 5 UT 95 dip in UT for 1 cycle 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 250 cycles 5 UT 95 dip in UT for 0 5 cycle 5 UT 95 dip in UT for 1 cycle 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 250 cycles Mains power quality should be that of a typical commercial or hospital environ...

Страница 22: ...on distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P1 2 d 2 P1 2 d 1 2 P1 2 80 MHz to 800 MHz d 2 3 P1 2 800 MHz to 2 7 GHz where P is the maximum output power rating of the transmitter In watts W according to the transmitter manufacturer and d Is the recommended separation distance in meters m Field strengths from fixed RF...

Страница 23: ...should be less than 3V m Table 4 Recommended separation distances between portable and mobile RF communications equipment and the model Fi P Recommended separation distances between portable and mobile RF communications equipment and the model Fi P The model Fi P is intended for use in electromagnetic environment in which radiated RF disturbances is controlled The customer or the user of the model...

Страница 24: ... the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 13 Statement Woodpecker reserves the right to change the design of the equipment the technique fittings instruction manual and the content of the original packing list at any t...