16
the wipe for at least 3 times.
• After each use, wipe the surface of the main unit and charging base
with a soft cloth soaked in clean water (distilled or deionized water) or a
clean disposable wipe. Repeat the wipe for at least 3 times.
8.12 Daily maintenance
1. When the device is not used, please turn off the power and unplug
the power supply plug.
2. If the Heating and Packing Instrument is in a low battery state for a
long time, the service life of battery will be shortened. Please charge it in
time if the battery level is low.
3. When the device is not used, please charge it for 1 hour once a
month.
8.13 Repair of device
This product does not contain self-repairing spare parts. If there
is any abnormality in the equipment, please contact our company
for maintenance and do not disassemble without authorization. With
our company's consent, we will provide circuit diagrams, component
part lists, descriptions, calibration instructions to assist to SERVICE
PERSONNEL in parts repair.
9 Troubleshooting
Fault
Cause
Solution
No indications,
no response
1. Inadequate battery power
2. Battery is damaged.
3. The charging interface is
short-circuited, causing the
lithium battery to enter a
protection state;
4. Heating and Packing
Instrument is damaged.
1. Connect to power
supply to charge. /
Replace the battery.
2. Replace the battery.
3. Remove the substance
that causes the short
circuit, put the device
into the charging base
to charge, and then the
device will return to
normal;
4. Contact local distributor
or manufacturer.
Содержание 28-780