10
8. Do not install, remove, or replace the Work Tip under heating state.
Please power off before replace the Work Tip.
9. The Work Tip must be correctly installed to prevent it from falling
off.
10. When the Work Tip is bent or worn, it will cause uneven heating.
The operator should replace the Work Tip in time according to the clinical
conditions;
11. After operation, please turn off the power immediately.
Woodpecker is specialized in producing medical instrument. We are
only responsible for the safety on the following conditions:
a) The maintenance, repair, and modification are made by the
manufacturer or the authorized dealers.
b) The charged components are original of “Woodpecker” and
operated according to instruction manual.
8 Cleaning, Disinfection, Sterilization and
Maintenance
The cleaning, disinfection and sterilization of Work Tip and Work Tip
Protector are as follow Unless otherwise stated, they will be hereinafter
referred to as “products”.
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid
pH <5) will reduce the life span of products. And in such cases, the
manufacturer takes no responsibility. This device shall not be exposed to
high temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization
cycles. The materials used in manufacture were selected accordingly.
However with every renewed preparation for use, thermal and chemical
stresses will result in ageing of the products. The maximum number of
sterilizations for Work Tip and Work Tip Protector is 300 times.
8.1 Initial processing
8.1.1 Processing principles
It is only possible to carry out effective sterilization after the
completion of effective cleaning and disinfection. Please ensure that, as
part of your responsibility for the sterility of products during use, only
sufficiently validated equipment and product-specific procedures are