24
Low level input voltage: 0 to +0,5 V
Hi level input voltage: +3,0 to +30 V
Auxiliary power U: 9 to 30 V DC (depends on voltage connected to DTR and/or
RTS minus about 1V), overall current max. 100 mA
Cable for input signal wiring: shielded, max. 10 m long, cable should not be led in parallel
along power cabling.
EMC:
EN 61326-1, EN 55011
10.1.2.
AR006684
- temperature and humidity regulator
Temperature:
Accuracy:
± 0,4 °C (±0,7 °F)
Range:
-30 to +105 °C (-22 to 221 °F), includes probe cable
Resolution: 0,1 °C (0,2 °F)
Relative humidity:
Accuracy:
± 2,5 %RH from 5 to 95 %RH at 23 °C (73,4 °F)
Range:
0 to 100 %RH
Resolution: 0,1 %RH
Measuring temperature and humidity range is limited in accordance with the graph below!
Response time
with stainless steel mesh sensor cover (F5200B) and bronze sensor cover (F0000
-
selectable option), air flow approximately 1 m/s:
temperature: AR006684, t90 < 6 min (temperature step 20 °C (36 °F))
relative humidity: t90 < 30 s (humidity step 65 %RH, constant temperature)
10.1.3.
AR006683
- temperature regulator, external probe
designed for two wires connection of external Pt1000 temperature sensor
Probe:
Pt1000/3850 ppm connected by shielded cable of maximum length 10m
Range:
-200 to +600 °C (-328 °F to 998.6 °F) Resolution:
0,1 °C (0,2 °F)
Accuracy:
±0,2 °C (±0,4 °F) without probe
3
There is possible to change response time. For more details see file “Description of calibration and adjustment
procedure.pdf”, which is installed together with Users software.