coMMissioning the gtl-3 test Device
12
ADICOS GTL-3
430-2410-102 EN11
6 Commissioning the GTL-3 test device
NOTICE!
The ADICOS GTL-3 tester works with controlled combustion, which can cause
fires and explosions if used improperly.
• Do not open the housing within hazardous areas
• Only operate the unit when it is approved for hot work!
• Only operate the unit with a glow filter!
• Only operate the device with the GTL-3 housing firmly closed!
CAUTION!
During operation of the ADICOS GTL-3 harmful vapours are generated.
• Do not inhale vapours!
Tips and recommendations
Before starting up the ADICOS GTL-3 make sure that:
• The batteries of the SOLO battery bar are sufficiently charged!
• A new smouldering rod in the holder GTL-3 smouldering rod is inserted
and screwed together!
• The fan unit is firmly screwed to the GTL-3 housing
• A new embers filter is inserted into the GTL-3 gas funnel and properly
clamped!
6.1 Ignite smouldering sticks
1
Remove the GTL-3 protective tube
2
Light smouldering rods at the upper end
3
Replace the GTL-3 protective tube, guiding the re-
cess of the protective tube over the anti-twist pro-
tection on the fan unit
4
Mount the GTL-3 housing search unit (–› Chap. 6.2,
Mounting the test head)