![Greyline Portaflow PT400 Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/greyline/portaflow-pt400/portaflow-pt400_user-manual_2264647042.webp)
Page
42
www.greyline.com
Appendix A:
Specifications
GENERAL
NEW! DSP Measurement
Technique:
Transit time.
Timing Resolution:
50 pico-second, continuous signal level indication on display.
Improved! Flow Velocity
Range:
Minimum Velocity 0.1m/s; Max Velocity 20m/s: Bi-directional.
Turn Down Ratio:
100:1
Accuracy:
±0.5% to ±2% of flow reading for flow rate >0.2m/s and Pipe ID >75mm.
±3% of flow reading for flow rate >0.2m/s and Pipe ID in range 13mm - 75mm.
±6% of flow reading for flow rate < 0.2m/s.
Repeatability:
±0.5% of measured value or ±0.02m/s whichever is the greater.
NEW! Reynolds Number
Correction:
Flow velocity corrected for Reynolds number over entire velocity range.
Response Time:
< 400ms depending on pipe diameter.
Selectable Flow Units:
VELOCITY: m/sec, ft/sec.
VOLUME: l/s, l/min, l/h, gal/min, gal/h, USgals/min, USgals/h, Barrel/h, Barrel/day, m³/s,
m³/min, m³/h.
Selectable Volume Units:
l, gal, USgals, Barrel, m³.
Total Volume:
12 Digits - forward and reverse.
APPLICABLE FLUID TYPES
Fluid Condition:
Clean liquids or oils that have less than 3% by volume of particulate content. Applications include river water,
sea water, potable water, demineralised water, glycol/water mix, hydraulic systems and diesel oil.
APPLICABLE PIPE TYPES
Pipe Materials:
Any sonic conducting medium such as Carbon Steel, Stainless Steel, Copper, UPVC, PVDF, Concrete,
Galvanised Steel, Mild Steel, Glass, Brass. Including Lined Pipes - Epoxy, Rubber, Steel, Plastic.
Pipe Dimension (outside
diameter):
Min 13mm; Max 4000mm with D sensor set.
Pipe Wall Thickness:
1mm - 75mm.
Pipe Lining:
Applicable pipe linings include Rubber, Glass, Concrete, Epoxy, Steel.
Pipe Lining Thickness:
0mm – 10mm.
Pipe Wall Temperature
Range:
Standard sensor operating temperature is -4°F to 275°F (-20°C to +135°C).
Optional high temperature sensor operating temperature is -4°F to 392°F ( -20°C to +200°C).
TRANSDUCER SETS
Standard:
Temperature Range -4°F to 275°F (-20°C to +135°C)
PF400A – Type 'A-ST' (2MHz)
PF400B – Type 'B-ST' (1MHz)
Содержание Portaflow PT400
Страница 1: ... PORTAFLOW PT400 Portable Ultrasonic Flowmeter User s Guide www greyline com ...
Страница 45: ...Page 45 www greyline com PIPE CHARTS ...
Страница 46: ...Page 46 www greyline com ...
Страница 47: ...Page 47 www greyline com ...