Temperature range
Interpretation:
You have entered an out-of-range value for the fluid Temperature.
Accepted temperature range -4°F to 572°F (-20°C to +300°C).
Response:
Enter a valid number.
Invalid Date or Time
Interpretation:
The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.
Response:
Enter a valid Date and Time.
Sensors: INVALID
Interpretation:
The selected temperature is higher than the maximum allowed for
the sensor type.
Response:
Select alternative sensors or change the temperature.
Mode: Err Typ
Interpretation:
The selected sensors are invalid and the mode cannot be verified.
Response:
Select a valid sensor type and choose a mode that gives a non-zero
separation distance.