Page
12
www.greyline.com
3
On each guide rail, attach one end of a securing chain to a hook on the tensioning bar (B), wrap the chain
around the pipe (G) and then attach it to the hook on the other end of the tensioning bar whilst keeping the
chain as tight as possible.
4.
Rotate the complete guide rail assembly so that it is approximately 45° with respect to the top of
the pipe. Then tighten the chain by turning the tensioning thumb-wheel (A) on each guide block until the
assembly is securely attached to the pipe.
Note:
If you are unable to get sufficient tension on the chain to hold the assembly in place, fully slacken the
tensioning thumb-wheel and shorten the effective length of the chain wrapped around the pipe by connecting
the tensioning bar to the next link in the chain, then re-tension.
2.2.3
Mounting the transducers
Figure 2.3 Mounting the transducers
1.
Slide the transducer cover plate (A) fully towards the outside of the guide assembly to allow
sufficient access to fit the transducer.
2.
Clean the face of the transducer, removing all traces of dirt and grease.
3.
Apply a 1/8” (3mm) bead of ultrasonic couplant along the centre length of the transducer (E).
4.
Fit the transducer into the guide block – ensuring the lugs on the sides of the transducer are
correctly located into the slots on the sides of the guide block (B).
5.
Slide the transducer cover plate (A) over the top of the transducer and tighten the thumbscrew
(C) finger tight to secure the transducer in place. When securing the cover plate take care to leave sufficient
room around the transducer connector (D) to connect the cable.
6.
Repeat the above steps for the second transducer.
7.
Connect the transducers to the Portaflow instrument using the coaxial cables provided. The RED
cable must be connected to the upstream transducer and the BLUE cable to the downstream transducer. If
you observe negative flow, swap the red and blue cables at the sensors.
C: Transducer cover plate securing screw.
B: Transducer locating slot/lug.
D: Transducer cable connection.
A: Transducer cover plate.
G: Chain.
E: Ultrasonic
couplant
application.