Page 18
TTFM 1.0 Transit Time Flow Meter
DATA LOGGING (OPTIONAL)
Refer to Options section of this manual.
SPECIAL FUNCTIONS
Language
Select
English
,
French
or
Spanish
Analog Out
Select 4-20mA or 0-5V mode for the analogue
output.
Backlight
Select
High
,
Medium
or
Low
for continuous
backlight.
Select
Key Hi/Lo
for high backlight for 1 minute
after a keypress and then
Lo
backlight until a key
is pressed again.
Select
Key High
,
Med
or
Low
for backlight for
1 minute after a keypress and then backlight off until
a key is pressed again.
Reset Totalizer
Press
and select
Yes
to erase and restart the
totalizer at zero.
Negative Totals
Select
Yes
to have reverse flow readings deducted
from the totalizer. Select
No
to totalize forward
flow only and ignore reverse flow.
Cal Constant
Factory set during calibration. (Refer to the
calibration certificate supplied with your instrument.)
Restore Defaults
Select
Yes
and press
to erase all user settings
and return the instrument to factory default settings.
New Password
Select any number from 0000 to 9999 and press
.
Default setting of 0000 will allow direct access to the
calibration menus. Setting of any password greater
than 0000 will require the password to be entered to
access the calibration menus.
Press
to return to
Menu Selections
.
Содержание TTFM 1.0
Страница 2: ...Note This page has been left blank intentionally ...
Страница 4: ...Page 4 TTFM 1 0 Transit Time Flow Meter SENSOR INSTALLATION ...
Страница 6: ...Page 6 TTFM 1 0 Transit Time Flow Meter CONNECTIONS ...
Страница 9: ...Page 9 TTFM 1 0 Transit Time Flow Meter CALIBRATION MENU ...
Страница 10: ...Page 10 TTFM 1 0 Transit Time Flow Meter ICONS ...
Страница 24: ...Page 24 TTFM 1 0 Transit Time Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS BAD GOOD ...
Страница 36: ...Page 36 TTFM 1 0 Transit Time Flow Meter SENSOR INTRINSIC SAFETY ...