Page 28
TTFM 1.0 Transit Time Flow Meter
SENSOR CABLE RESISTANCE TEST
Unplug the green sensor terminal from the Transit Time board and connect the sensor wires. With a
multimeter, perform resistance checks for each set of wires. One single loose terminal may cause false
readings.
Test across shield and core of each wire: TDR1 (black) and TDR2 (black/white). Resistance should be
around 10K ohms for any cable length. High readings indicate an open circuit and low readings indicate
a short or partial short in the sensor cable connections.
Note: The TTFM 1.0 will automatically detect connectivity to the sensors. Confirm that TTFM 1.0
indicates “
Sensor
Good
’ in the messages menu.