Page 15
TTFM 1.0 Transit Time Flow Meter
Crossings
1 = Z mounting
2 = V mounting
4 = W mounting
Zero Tare
To suppress readings or fluctuations at zero flow. Set
Calibration
/
Damping
to
5%
and under no flow conditions
and full pipe select
Yes
to force readings to zero.
Sens Space
Displays the calculated sensor spacing
Velocity
Displays the measured velocity.
Signal Strength
Displays magnitude of signal being received by the
ultrasonic sensor.
Press
to return to Menu Selections.
NOTE:
Sensor separation distance is automatically calculated by
the instrument and will be displayed in the
Setup
menu