TFT32 CALIBRATION MENU
PRESSURE SENSOR INPUT
TFT32 Tank Farm Transmitter
Manual Series 2.4
Page 8
LO
1056
HI
1.000
dp10
98.0
100.0
ro FF
r on
r o c 1
100=
999.9
0.0
0=
123.4
CALU
CM
CM
PERCENT
PERCENT
CM
CM
CM
CM
CM
LEVEL CM
PERCENT INCHES CM
mA DISPLAY
Displays mA signal as received from sensor
or
to increase or
decrease.
IN-SITU CALIBRATION
Simulate 4mA with 4-20mA input device and enter the minimum
value of input device.
With level 70% of tank enter the current level.
Simulate 20mA with 4-20mA input device and enter the
maximum value from the input device.
Ç
È
>
NORMAL MODE
CALIBRATION UNITS SELECTION
Press
to select inches, cm, percent or blank (other units)
OUTPUT SPAN CALIBRATION
"0=" Enter minimum reading to be displayed by the TFT32
"100=" Enter maximum reading to be displayed by the TFT32
Wait for the numeric value then press
RELAY OPERATION CHOICE.
Press
to select:
3 = Signal Loss Only
1 = Level Alarm
4 = Always ON
2 = Level + Signal Loss
5 = Always OFF
RELAY SET POINTS
Wait for numeric value and press
and
to adjust Set Points
(in percentage of Span)
DAMPING (0-15 secs)
Press
to adjust
TRANSMITTER/TANK ADDRESS
Press
to select 1-32
SERIAL OUTPUT BAUD RATE
Press
to select
19 = 19200
4 = 4800
9 = 9600
2 = 2400
OUTPUT SIMULATION
Wait for numeric value and press
and
to increase or
decrease Simulated Output (Automatically returns to NORMAL
mode in 10 minutes)
STORE YES?
Press
to store Calibration (displays -C.S-)
Press
Calibration Not Stored
Ç
Ç
Ç
È
Ç
Ç
Ç
Ç
È
Ç
È
YES?
-C .S-
100.0
OPSL
bd19
Ad 1
YES
NO
3,4,5, 6
1,2
18
9
9
10
11
15
16
16
17
See Page
#
20.0