LOW LEVEL CALIBRATION (IN-TANK)
With tank level <30% of maximum level (
100=
) wait for the numeric value and press
Ç
to enter the
current tank level.
HIGH LEVEL CALIBRATION (IN-TANK)
With tank level >70% of maximum level (
100=
) wait for the numeric value and press
Ç
to enter the
current tank level.
Note: the greater the range between LO and HI levels for this procedure, the better the accuracy of your
calibration. (eg: LO at 10% and HI at 90% will produce a more accurate calibration than LO at 40% and
HI at 60%).
Note: If only one value is changed during a calibration then the second calibration point remains the
unchanged value from the last calibration (ie. You can perform a single point calibration to improve
upon the previous calibration).
Note: Sensor input difference between HI and LO must be greater than 2mA.
TFT32 Tank Farm Transmitter
Manual Series 2.4
Page 11