Page 25
PTFM 1.0 Portable Transit Time Flow Meter
Possible Causes:
Corrective Action:
Coupling compound washed out, or sensor loose on
pipe.
Remount sensor
Use Dow Corning Silicone #4
METER READING HIGHER THAN EXPECTED?
Calibration Error
Review calibration menu. Pipe dimensions and
fluid selection/fluid velocity.
Higher flow rate than expected
Investigate pump/valves. Compare velocity
with alternate instrument.
Erratic measurement (set damping to 0% to check) due
to electrical noise or poor signal quality.
Try adjusting sensor spacing (+/- 10%) and
contact Greyline for further assistance.
Pipe not Full
Verify pipe is full by mounting sensors at top
of pipe and check echo icon. No echo if pipe is
not full.
High viscosity fluid
Laminar flow profile due to high viscosity fluid
requires an adjustment to Cal Const.
BATTERY CONDITIONING
To restore or condition your PTFM 1.0 battery
for best performance fully discharge the unit
(leave it on with backlight lit until it switches
off) and then charge for 6 hours, disconnect
charger momentarily, reconnect and charge
again for 6 hours.