Page 15
PTFM 1.0 Portable Transit Time Flow Meter
CALIBRATION
Press the
to
Calibration
and
to enter. Use
or
to position
before
each menu item and
to enter. When settings are completed press
to store and
return to the
Calibration
menu.
*20mA at
Press
then
or
to change the numbers and decimal point.
Use this menu to set the corresponding flow rate that will be
represented by 20mA analog output. If maximum flow is
unknown, enter an estimated flow rate and observe actual flow to
determine the correct maximum value. Any velocity or flow rate
up to +40 ft/sec (12.2 m/sec) may be selected.
*4mA at
Press
or
to set the flow rate corresponding to 4mA analog
output. This setting may be left at zero flow (or velocity or can
be raised to any value less than the 20mA setting, or lowered to
any velocity or corresponding flow rate down to -40 ft/sec (-12.2
m/sec).
Min Flow
Flow rates below this setting will be displayed as zero flow.
Damping
Increase damping to stabilize readings under turbulent flow
conditions. Decrease for fast response to small changes in flow.
Damping is shown in percentage (maximum is 99%). Factory
default is 5%.
Press
from the
Units/Mode
display to return to
Menu Selections
.
*
Note
4-20mA circuitry is only powered by the AC power module.
To conserve power this output is not active in battery power mode.
Содержание PTFM 1.0
Страница 2: ...Page 2 Note This page has been left blank intentionally ...
Страница 6: ...Page 6 PTFM 1 0 Portable Transit Time Flow Meter SENSOR INSTALLATION ...
Страница 8: ...Page 8 PTFM 1 0 Portable Transit Time Flow Meter CALIBRATION MENU ...
Страница 23: ...Page 23 PTFM 1 0 Portable Transit Time Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS BAD GOOD ...
Страница 45: ......