Page 27
PTFM 1.0 Portable Transit Time Flow Meter
What is the purpose of the Signal Strength Display?
The primary function of the signal strength display is to assist as a feedback when mounting sensors.
Signal Strength can also be a useful diagnostics tool when troubleshooting problems with an installation.
A low signal strength (< 10%) will cause the PTFM to be more susceptible to environmental noise and
may indicate a problem with the installation or other qualitative issues.
Does the PTFM 1.0 require periodic recalibration?
PTFM 1.0 calibration does not normally drift over time. Greyline offers a calibration service to verify
instrument accuracy.
ISO 9000 or similar quality management systems may require periodic and verifiable recalibration of
flow meters. PTFM 1.0 Flow Meters may be returned to Greyline for factory calibration and issue of a
new NIST traceable certificate. Refer to the ‘Product Return Procedure’ section of this manual for return
instructions.
Can the internal batteries be replaced?
The built-in rechargeable NiMH battery pack is not user-serviceable. The meter should be returned to
Greyline for battery service.