Page 16
PTFM 1.0 Portable Transit Time Flow Meter
DATA LOGGING
Setup
Select
Data Logging
from Menu Selections.
Log Site ID
Enter a number from 00 to 99. The site ID will become part of
the downloaded file name to help distinguish downloads from
different instruments. Press
to store the setting.
Mode
Select Velocity (e.g. ft/sec or m/sec).
Flow
(e.g. USGPM or
l/sec). Press
to store the setting.
Set Date
Press
or
to scroll and select Month, Day and Year. Press
to store the setting.
Set Time
Press
or
to select the current time in Hours, Minutes and
Seconds. Press
to store the setting.
Interval
Press
or
to select the logging interval.
Press
to store the setting.
Log
Select
Start
then
to start logger. Select
Stop
then
to stop
logger. Select
Delete
then
to clear all previous logging
sessions.
RETRIEVE LOG FILE
Install Greyline Logger on your PC or laptop. Refer to the Help menu in the
program for detailed instructions.
- Connect the PTFM 1.0 to the PC using the supplied USB cable.
- Install the USB driver program from the install CD.
- Start the Greyline Logger Software.
- Select "xxxx scan for USB instruments xxxx" in the drop down window at the
top of the main window. PTFM 1.0 will be indicated.
- Click the download icon to start transferring data.
- Downloaded data appears in a pop-up window.
Содержание PTFM 1.0
Страница 2: ...Page 2 Note This page has been left blank intentionally ...
Страница 6: ...Page 6 PTFM 1 0 Portable Transit Time Flow Meter SENSOR INSTALLATION ...
Страница 8: ...Page 8 PTFM 1 0 Portable Transit Time Flow Meter CALIBRATION MENU ...
Страница 23: ...Page 23 PTFM 1 0 Portable Transit Time Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS BAD GOOD ...
Страница 45: ......