APPENDIX A – OPTIONS
EXTRA SENSOR CABLE
(OPTION DXC)
Each Greyline flow meter includes 20 ft / 6m (or 50 ft / 15 m optional) continuous shielded coaxial pair
cable. Additional cable and Cable Junction Box (Option DJB) may be ordered with the Flow Meter, or
the cable may be spliced and extended up to 500 ft (152m) as required during installation. No
adjustment is required when the sensor cable is extended or shortened. Use only Greyline shielded
coaxial pair (RG174U) cable.
Extended sensor cable should be installed in conduit for mechanical protection. Recommended
installation with a metal junction box (Option DJB) is illustrated below:
SENSOR CABLE JUNCTION BOX
(OPTION DJB)
Optional Watertight steel NEMA4 Junction Boxes with terminal strips are available from Greyline
Instruments.
Page 27
DFM-IV Doppler Flow Meter
Manual Series A.6
DIMENSIONS
OPTION DJB - JUNCTION BOX