4.2  „Func-CYCL“: Automatic Recording With Selectable Logger-Cycle-Time

The Logger-Cycle-Time is settable (p.r.t. Configuration). For example „CYCL“ = 60: A measuring is recorded
after each 60 seconds.
When the slow measurement "rAtE-Slo" is chosen, additionally a low power function is available: „Lo.Po“.
If „Lo.Po“ is on, the device only will take a measurement at the point of time of the recording. In between the
recordings the measuring shut’s down. This decreases the power consumption enormously and therefore is
recommended e.g. for long time recordings where no mains adapter is available.

Max. number of measurings:

9999

Cycle time:

1...3600 seconds (=1h), selectable in the configuration

A measuring contains:

!

  rAtE SLo:

- current measuring value at the time of recording
- min peak, max peak since the last recording

!

  rAtE FASt,P.dEt

- arithmetic mean value since the last recording
- min peak, max peak since the last recording

Starting a recording:

By pressing "Store" (key 6) for 2 seconds the recording will be initiated. After that the display shows
‘St.XXXX‘ for a short time whenever a measuring is recorded. XXXX is the number of the measuring 1..9999.

If the logger memory is full, the display will show: 

 

 The recording automatically will be stopped.

If Low-Power-Logger-Function „Lo.Po = on“ the device switches itself off as soon as the memory gets filled.

Stopping the recording manually:

By pressing "Store" (key 6) the recording can be stopped manually. Then the following choice appears:

 

The selection can be made by 

5

 (key 2) and 

6

 (key 5). "Quit" (key 6) enters the choice.

Note: 

If you try to switch off the instrument in the cyclic recording operation You will be asked
once again if the recording is to be stopped.
The device can only be switched off after the recording has been stopped!
The Auto-Power-Off-function is deactivated during recording!

Clear Recordings:

When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory will be displayed:

The selection can be made by 

5

 (key 2) and 

6

 (key 5). "Quit" (key 6) enters the choice.

Stop the
recording

Do not stop the
recording

Clear all
recordings

Clear nothing
(cancel menu)

Содержание GMH3150

Страница 1: ...H31 0 31 6C 06 User s Manual Handheld Digital Pressure Meter GMH3150 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...nual will be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those s...

Страница 3: ...may not be detected correctly When connecting the sensor the connector may not lock correctly In such case take the plug not at the casing but at the buckling protection at the end of the plug If plu...

Страница 4: ...eaning Menu 4 4 4 4 5 or 6 Set Configuration Generic Configurations Unit mbar bar Unit Unit of display SL oFF on Sea level correction on or off Alti 2000 9999 Altitude Input of altitude above sea leve...

Страница 5: ...be measured the pressure loss resulting from the actual level above sea level has to be considered To correct the measuring display activate the Sea Level Function SL p r t chapter 2 Then enter the al...

Страница 6: ...g the point of time when they were measured Please check the settings when necessary If the battery was replaced the refering menu CLOC will automatically be started 3 Measuring Of Water Level Display...

Страница 7: ...ime when store key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is d...

Страница 8: ...ak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time whenever a measuri...

Страница 9: ...base address 2 min peak p r t chapter 4 3 max peak p r t chapter 4 Note The measuring alarm and display range values read back from the interface are always in the selected measurement unit mbar bar S...

Страница 10: ...steel pressure sensors types GMSD MRE GMSD BRE GMSD BAE For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 7 Error An...

Страница 11: ...tivated Min Max Alarm The measuring value is constantly monitored for the min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and...

Страница 12: ...Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1 mm tubes Housing ABS with mounting eyelet...

Отзывы: