GREISINGER GMH3150 Скачать руководство пользователя страница 5

H31.0.31.6C-06

User’s Manual GMH3150

page 5

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

  

2.1  Different Kinds Of Measuring: „rAtE-Slo, -P.dEt, -FASt“

Three different kinds of measuring pressure are supported. Two of them are working with high measuring
frequency of more than 100 measurings per second. If one of them was chosen in the configuration (see
above), this will be displayed in the secondary display: „P.dEt“ or „FASt“.

2.1.1 rAtE-Slo:

Standard 

Measuring

Measuring rate 4Hz, averaging and filter functions are active.
Application: Measuring of slowly changing or static pressures, e.g.  measuring of leakproofness, atmospheric
pressure...
Highest accuracy, high noise immunity (EMI and unstable measuring signals), low power consumption.

2.1.2 rAtE-P.dEt: Peak 

detection

Measuring rate >100Hz, the value is displayed unfiltered.
Application with logger function: Measuring of short pressure peaks or fast changing pressures with a
resolution of < 10ms. The cyclic logger function records the arithmetic mean value, the highest and the
lowest peak of the referring time interval.
Attention: higher power consumption, measuring is sensitive to noise (EMI,..).

2.1.3  rAtE-FASt:

Fast filtered measuring

Measuring rate >100Hz, the value is filtered slightly (higher noise immunity than P.dEt, small peaks will be
filtered out), apart from that identical behaviour like P.dEt.

2.2  Sea Level Correction For Absolute Pressure Sensors

The device displays the absolute pressure measured at the sensor. This is not necessarily the same like the
values given by weather stations! The weather stations‘ values are giving the pressure at sea level. Usually
the sensor is placed above sea level and therefore, if the value at sea level(zero) is to be measured, the
pressure loss resulting from the actual level above sea level has to be considered! To correct the measuring
display activate the „Sea-Level-Function“ (SL, p.r.t. chapter 2). Then enter the altitude above sea level of the
sensor‘s location in meters (Alti, p.r.t. chapter 2). When activated, the display shows the SL-arrow and the
device displays the pressure value at sea level.

2.3 Averaging Function

The averaging function concerns the display values (LCD and interface). It is completely independent from
the averaging of the logger function, please don’t mix them up!
The averaging integrates the measuring values during a selectable period of time and then calculates the
average display value. It is independent from the selected kind of measuring (slow, fast, peak detect)

 .

As long as not enough values are collected (selected averaging time) to calculate a average value, the upper
display shows “----“, the lower display a ‘countdown‘.
During an active low-power-logging procedure the averring is always deactivated
Function of min/max-value memory during averaging:
-  If averaging is activated and slow measuring is selected (rAtE-Slo), the min-/max-value memory refers to

the average display value.

-  If averaging is activated and fast measuring is selected (rAtE-FASt or P.dEt) , the min-/max-value memory

refers to the internal measured values (fast peaks can be detected).

2.4  Power Off Time

If there won’t be pressed any key and no interface communication takes place for the time of the power off
time setting (P.Off), the device will be switched off automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.

2.5 Address

Up to 10 devices of the GMH3xxx- handheld-family can be connected to a serial interface at once
(depending on interface converter, e.g. GRS3105: 5 devices). To get access to each device the base
addresses of the devices have to be different. For example choose 01 for the first, 11 for the second device
and so on. See also chapter 5.

Содержание GMH3150

Страница 1: ...H31 0 31 6C 06 User s Manual Handheld Digital Pressure Meter GMH3150 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...nual will be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those s...

Страница 3: ...may not be detected correctly When connecting the sensor the connector may not lock correctly In such case take the plug not at the casing but at the buckling protection at the end of the plug If plu...

Страница 4: ...eaning Menu 4 4 4 4 5 or 6 Set Configuration Generic Configurations Unit mbar bar Unit Unit of display SL oFF on Sea level correction on or off Alti 2000 9999 Altitude Input of altitude above sea leve...

Страница 5: ...be measured the pressure loss resulting from the actual level above sea level has to be considered To correct the measuring display activate the Sea Level Function SL p r t chapter 2 Then enter the al...

Страница 6: ...g the point of time when they were measured Please check the settings when necessary If the battery was replaced the refering menu CLOC will automatically be started 3 Measuring Of Water Level Display...

Страница 7: ...ime when store key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is d...

Страница 8: ...ak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time whenever a measuri...

Страница 9: ...base address 2 min peak p r t chapter 4 3 max peak p r t chapter 4 Note The measuring alarm and display range values read back from the interface are always in the selected measurement unit mbar bar S...

Страница 10: ...steel pressure sensors types GMSD MRE GMSD BRE GMSD BAE For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 7 Error An...

Страница 11: ...tivated Min Max Alarm The measuring value is constantly monitored for the min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and...

Страница 12: ...Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1 mm tubes Housing ABS with mounting eyelet...

Отзывы: