4  Operation Of Logger

The device supports two different logger functions:

„Func-Stor“:

 

each time when „store“ (key 6) is pressed a measurement will be recorded.

„Func-CYCL“:

measurements will automatically be recorded each interval, which was set in the logger
menu ‚CYCL‘ until the logger will be stopped or the logger memory is full. The recording is
started by pressing „Store“ 2 seconds.

The logger records 3 measurement results each time:

current or mean value (depending on logger setting, see below), min peak and max peak.

Min and max peak

 are the minimum resp. the maximum of the measured values since the last recording.

Using them allows f.e. analysis of fluctuating  pressures. For the evaluation of the data the software
GSOFT3050 has to be used. The software also allows easy configuration and starting of the logger.
When the logger is activated (Func Stor or Func CYCL) the hold function is no more available, the key 6 is
solely used for the operation of the logger functions.

4.1  „Func-Stor“: Storing Single Measurements

Each time when „store“ (key 6) is pressed a measurement and its time stamp will be recorded.
The recorded data can be viewed either in the display (when calling the configuration an additional menu
„REAd LoGG“ is displayed, see below) or by means of the interface and a PC with GSOFT3050-software.

Max. number of measurings:

99

A measuring contains:

 

- current measuring value at the time of recording
- min peak, max peak since the last recording
- time and date of the recording

After each recording „St. XX“ will be displayed for a short time. XX represents the number of the recording.

If logger memory contains recordings already:

When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory will be displayed:

The selection can be made by 

5

 (key 2) and 

6

 (key 5). "Quit" (key 6) enters the choice.

If the logger memory is full, the display will show: 

 

Viewing Recorded Measurings

Within the „LoGG Stor“ function the measurings can be viewed directly in the display not only by means of a
computer (like at „Func CYCL“): press 2 seconds „Set“ (key 4): The first menu displayed now is „rEAd LoGG“
(read logger data). After pressing 

4

 (key 3)  the measurement recorded last will be displayed, changing

between the different values referring to the measurement also is done by pressing 

4

.

Changing the measurement is done by pressing the keys 

5

 or 

6

.

Clear all
recordings

Clear the last
recording

Clear nothing
(cancel menu)

Содержание GMH3150

Страница 1: ...H31 0 31 6C 06 User s Manual Handheld Digital Pressure Meter GMH3150 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...nual will be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those s...

Страница 3: ...may not be detected correctly When connecting the sensor the connector may not lock correctly In such case take the plug not at the casing but at the buckling protection at the end of the plug If plu...

Страница 4: ...eaning Menu 4 4 4 4 5 or 6 Set Configuration Generic Configurations Unit mbar bar Unit Unit of display SL oFF on Sea level correction on or off Alti 2000 9999 Altitude Input of altitude above sea leve...

Страница 5: ...be measured the pressure loss resulting from the actual level above sea level has to be considered To correct the measuring display activate the Sea Level Function SL p r t chapter 2 Then enter the al...

Страница 6: ...g the point of time when they were measured Please check the settings when necessary If the battery was replaced the refering menu CLOC will automatically be started 3 Measuring Of Water Level Display...

Страница 7: ...ime when store key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is d...

Страница 8: ...ak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time whenever a measuri...

Страница 9: ...base address 2 min peak p r t chapter 4 3 max peak p r t chapter 4 Note The measuring alarm and display range values read back from the interface are always in the selected measurement unit mbar bar S...

Страница 10: ...steel pressure sensors types GMSD MRE GMSD BRE GMSD BAE For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 7 Error An...

Страница 11: ...tivated Min Max Alarm The measuring value is constantly monitored for the min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and...

Страница 12: ...Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1 mm tubes Housing ABS with mounting eyelet...

Отзывы: