GREISINGER GMH3150 Скачать руководство пользователя страница 4

H31.0.31.6C-06

User’s Manual GMH3150

page 4

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

  

Zero-Point Adjustment: 

(for rel. pressure sensors only) If there is no pressure applied to the pressure ports

the device will display 0. If  there is a permanent deviation (and device is operated under
steady conditions), a permanent zero point adjustment can be carried out.
To carry out the adjustment press button 3 for approx. 5 seconds.
(Please note: A zero-point adjustment can only be carried out if  the difference between the
value on display and the value calibrated on site is less than 2%! E.g. for the measuring
range of –1.00 ..+25.00mbar, =>zero-point adjustment up to 0.50mbar possible)
To recall the manufacturer`s calibration press button 3 for approx. 7 seconds.

Note: If a zero-point adjustment was carried out, this will be signalled by the short
displaying of „NuLL Corr“ when switching on the device.

2 Configuration

To change device settings, press 

Menu 

(key 4) for 2 seconds. This will call the configuration menu (main

display:  „SEt“).
Pressing key 

Menu

 changes between the menues, pressing 

4

(key 3) jumps to the referring parameters,

which can be selected with key 

4

(key 3).

The parameters can be changed with 

5

 (key 2) or 

6

 (key 5).

Pressing 

Menu 

again jumps back to the main configuration menu and saves the settings.

Quit

 (key 6) finishes the configuration and returns to standard measuring operation.

Menu

Param.

Values

Meaning

‚Menu‘

4

4

4

4

5

 or 

6

Set Configuration: Generic Configurations

Unit

mbar,bar..

Unit

: Unit of display

*

SL

oFF/on

Sea level correction

: on or off

*

Alti

-2000..9999

Altitude

: Input of altitude above sea level [m] (only if SL on)

*

Rate

: Measuring rate (p.r.t. chapter 2.1)

*

Slo

Slow

 measuring rate (4Hz filtered, low power consumption)

*

FASt

Fast

 measuring rate, filtered (>100Hz)

*

rAtE

P.dEt

Peak detection

: fast measuring rate, unfiltered (>100Hz)

*

1-120

Averaging period in seconds, used by the averaging function

t.AVG

oFF

Averaging function deactivated

1-120

Auto Power Off

 time in minutes

P.oFF

oFF

Auto Power Off

 

deactivated

SEt
ConF

Adr.

01,11..91

Base 

address

 of interface

Set Alarm: Settings Of Alarm Function

On

Alarm

 on, with horn-sound

no.So

Alarm on, without horn-sound

AL.

oFF

no alarm function

AL.Lo

Sensor-Min
... AL.Hi

Min alarm rail (not when AL. oFF, Sensor-Min is the lower display
range of connected sensor)

SEt
AL.

AL.Hi

AL.Lo ...
Sensor-Max

Max alarm rail (not when AL. oFF, Sensor-Max is the upper display
range of connected sensor)

Set Logger: Configuration Of Logger Function

*

CYCL

Cyclic

: logger function ‚cyclic logger‘

*

Stor

Store

: logger function ‚individual value logger‘

*

Func

oFF

no logger function

*

CYCL

1..3600

Cycle time of cyclic logger [seconds]

*

SEt
LoGG

Lo.Po

on/oFF

Low-power logger

 with very low power consumption

(only for cyclic logger and slow measuring rate)

*

Set Clock: Setting Of Real Time Clock

CLOC

HH:MM

Clock

: Setting of time  hours:minutes

dAtE

TT.MM

Date

:                            day.month

SEt
CLOC

YEAr

YYYY

Year

Note: 

If the logger memory contains data already, the menues/parameters marked with (*) can
not be invoked! If these should be altered the logger memory has to be cleared before!
(key 6, p.r.t. chapter 4)

Содержание GMH3150

Страница 1: ...H31 0 31 6C 06 User s Manual Handheld Digital Pressure Meter GMH3150 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...nual will be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those s...

Страница 3: ...may not be detected correctly When connecting the sensor the connector may not lock correctly In such case take the plug not at the casing but at the buckling protection at the end of the plug If plu...

Страница 4: ...eaning Menu 4 4 4 4 5 or 6 Set Configuration Generic Configurations Unit mbar bar Unit Unit of display SL oFF on Sea level correction on or off Alti 2000 9999 Altitude Input of altitude above sea leve...

Страница 5: ...be measured the pressure loss resulting from the actual level above sea level has to be considered To correct the measuring display activate the Sea Level Function SL p r t chapter 2 Then enter the al...

Страница 6: ...g the point of time when they were measured Please check the settings when necessary If the battery was replaced the refering menu CLOC will automatically be started 3 Measuring Of Water Level Display...

Страница 7: ...ime when store key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is d...

Страница 8: ...ak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time whenever a measuri...

Страница 9: ...base address 2 min peak p r t chapter 4 3 max peak p r t chapter 4 Note The measuring alarm and display range values read back from the interface are always in the selected measurement unit mbar bar S...

Страница 10: ...steel pressure sensors types GMSD MRE GMSD BRE GMSD BAE For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 7 Error An...

Страница 11: ...tivated Min Max Alarm The measuring value is constantly monitored for the min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and...

Страница 12: ...Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1 mm tubes Housing ABS with mounting eyelet...