User’s Manual GMH3150

page 3

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

  

main display: 

shows measuring value.

!

: indicates weak battery or other warnings

secondary display: 

min-, max- or hold value

Logg: 

appears, if logger function is chosen,
flashes when logger is running

AL: 

(not at all devices) flashes, if alarm exists

Units: 

an arrow points to

the chosen measuring unit

Tara:

 appears if tara-

function is activated.

SL:

 appears if sea-level-

correction is activated

1.2  Operation And Maintenance Advice

 

Battery Operation

If

and ´bAt´ are shown in the secondary display the battery has been used up and needs to be

replaced. The device will, however, operate correctly for a certain amount of time. If ´bAt´ is shown in the
upper display the voltage is too low to operate the device; the battery has been completely used up.
The battery has to be taken out, when storing device above 50°C.

Please note: We recommend to take out battery if device is not used for a longer period of time!

 

Mains Operation With Power Supply
Warning:

 

When using a power supply please note that operating voltage has to be 10.5 to 12 V DC.

Do not apply overvoltage!! Cheap 12V-power supplies often have excessive no-load voltage.
We, therefore, recommend using regulated voltage power supplies. Trouble-free operation is guaranteed
by our power supply GNG10/3000.
Prior to connecting the power supply to the mains make sure that the operating voltage stated at the
power supply is identical to the mains voltage.

!

 

Connecting/Changing Sensors

Do not use insuitable sensors. Connecting other devices/sensors as specificated may cause a damage to
the instrument and device/sensor! Switch off device before changing the sensor.
Connect sensor before switching on the device, otherwise the sensor may not be detected correctly.
When connecting the sensor the connector may not lock correctly. In such case take the plug not at the
casing but at the buckling protection at the end of the plug. If plug is entered correctly, it will slide in
smoothly. To disconnect sensor do not pull at the cable but at the plug (to open locking mechanism).

1.3 Connections

Connection for pressure sensors of the GMSD-family (p.r.t. chapter 10)
Interface: Connection for el. isolated interface adapter (p.r.t. chapter 5)
The mains adapter socket is located at the left side of the device.

1.4 Display

1.5 Basic Operation

When switching on 

the device and the logger function is not off the time of the integrated clock will shortly

be displayed. If a 

zero point adjustment

 was carried out the display shows shortly „nuLL Corr“.

After changing the battery

 the clock-setting menu is activated automatically (‚CLOC‘). Check the clock and

adjust, if necessary (p.r.t. chapter 2).

On-/Off-Switch
Tara:

Calling of tara function, zero point adjustment

min/max:

Showing the min- resp. max-memory in sec. display

Store/Quit:

  Calling of hold function resp. calling of logger functions (p.r.t. chapter 4)

Set/Menu:

Calling of configuration

Max Memory:

Pressing ´max´ (key 2) shows the maximum of the measured values. Pressing it again
hides it. To clear the max memory press key ´max´ for >2 seconds.

Min Memory:

Pressing ´min´ (key 5) shows the minimum of the measured values. Pressing it again
hides it. To clear the min memory press key ´min´ for >2 seconds.

Hold Function:

By pressing ´Store/Quit´ (key 6) the last measuring value will be held in the secondary
display. Pressing it again hides it. (only when logger = ‚off‘).

Tare Function:

By pressing ´Tara´ (key 3) the display will be set to 0. All measurings from then on will be
displayed relatively to the set  tare value. When tara function is activated, the arrow "Tara"
appears in the display. To deactivate tare function press ´Tara´  for >2 seconds.

Please Note: Activating/deactivating tara clears the max- & min-memories.

1

2

3

4

5

6

ON

OFF

max

Tara

Store

min

Set

Menu

Quit

!

Содержание GMH3150

Страница 1: ...H31 0 31 6C 06 User s Manual Handheld Digital Pressure Meter GMH3150 Version 5 1 for GMSD Pressure Sensors...

Страница 2: ...nual will be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those s...

Страница 3: ...may not be detected correctly When connecting the sensor the connector may not lock correctly In such case take the plug not at the casing but at the buckling protection at the end of the plug If plu...

Страница 4: ...eaning Menu 4 4 4 4 5 or 6 Set Configuration Generic Configurations Unit mbar bar Unit Unit of display SL oFF on Sea level correction on or off Alti 2000 9999 Altitude Input of altitude above sea leve...

Страница 5: ...be measured the pressure loss resulting from the actual level above sea level has to be considered To correct the measuring display activate the Sea Level Function SL p r t chapter 2 Then enter the al...

Страница 6: ...g the point of time when they were measured Please check the settings when necessary If the battery was replaced the refering menu CLOC will automatically be started 3 Measuring Of Water Level Display...

Страница 7: ...ime when store key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is d...

Страница 8: ...ak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time whenever a measuri...

Страница 9: ...base address 2 min peak p r t chapter 4 3 max peak p r t chapter 4 Note The measuring alarm and display range values read back from the interface are always in the selected measurement unit mbar bar S...

Страница 10: ...steel pressure sensors types GMSD MRE GMSD BRE GMSD BAE For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 7 Error An...

Страница 11: ...tivated Min Max Alarm The measuring value is constantly monitored for the min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and...

Страница 12: ...Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1 mm tubes Housing ABS with mounting eyelet...

Отзывы: