H74.0.14.6C-05
Operating Manual GMH 3692
page 7 of 12
Menu
Parameter Value
Description
key Menu
key
key
or
p.r.t
SET
(ONF
Set Configuration: General configurations
[H 2
P 02
hPa
Oxygen partial pressure display in hPa
*
P 02
mmHg
Oxygen partial pressure display in mmHg
LCD.2
T
Second. display always temperature
P
Second. display always absolute pressure
P T
Second. display alternates between temperature and abs. pressure
U N,T
T
°C
All temperatures in degree Celsius (ex works setting)
°F
All temperatures in degree Fahrenheit
[AL.P
1-PT
Simple one point calibration at air
2-PT
2 point calibration at air and 0% (e.g. N
2
) or 100 %
3-PT
2 point calibration at air and 0% (e.g. N
2
) and 100 %
[. INT
1
…
365
Calibration reminder period (in days)
OFF
No calibration reminder
A VTO
HLD
ON
Auto measuring value identification Auto Hold (when logger = off)
OFF
Standard hold function on key press (when logger = off)
P.OFF
1
…
120
Power-off delay in minutes. Device will be automatically switched off as
soon as this time has elapsed if no key is pressed/no interface
communication takes place. (ex works setting 20min)
OFF
Automatic power-off function deactivated (continuous operation)
ADR
0 1
,
1 1
,
2 1
,
…
9 1
Base address for serial interface communication
(ex works setting 01)
SET
(ORR
Set Corr: Input adjustment
0FFS
°C or °F
-5.0
°C ..
5.0
°C
or.
-9.0
°F
.. 9.0
°F
The zero point of the temperature measuring is shifted for the entered
value.
This can be used to compensate sensor and instrument deviations
OFF
No zero adjustment for temperature measurement (=0.0°)
S[AL
°C or °F
-5.00 ... 5.00
%
The slope of the temperature measurement is corrected by this value.
This can be used to compensate sensor and instrument deviations
OFF
No slope adjustment for temperature measurement (=0.00)
0FFS
hPa
-20
..
20
hPa
The zero point of the pressure measuring is shifted for the entered value.
This can be used to compensate sensor deviations
OFF
No zero adjustment for pressure measurement (=0.0°)
SET
AL
Set Alarm: Einstellung der Alarmfunktion
AL. 1
ON
/
NO.SO
Messkanal Sauerstoff: Alarm an mit Hupe / Alarm an ohne Hupe
OFF
keine Alarmfunktion für Messkanal Sauerstoff
AL.,N
[ONC
Alarmkanal Sauerstoff: Konzentration in %
P. 02
Alarmkanal Sauerstoff: Partialdruck in hPa oder mmHg
A 1.LO
z
.B.
0.0..100.0
%
Min-Alarm-Grenze Sauerstoff (nicht bei AL. 1. oFF)
A 1.H,
z
.B.
0.0..100.0
%
Max-Alarm-Grenze Sauerstoff (nicht bei AL. 1. oFF)
AL. 2
ON
/
NO.SO
Alarm Temperaturmessung an mit Hupe / Alarm an ohne Hupe
OFF
keine Alarmfunktion für Temperaturmessung
A2.LO
-5.0 ..
+
50.0
°C
Min-Alarm-Grenze Temperatur (nicht bei AL. 2. oFF)
A2.H,
-5.0 ..
+
50.0
°C
Max-Alarm-Grenze Temperatur (nicht bei AL. 2. oFF)
P
ressing “menu” and “store” at the same time for more than 2 seconds will reset the device to
factory defaults
If no key is pressed within 2 minutes the configuration will be aborted. All changes will not be saved!