GREISINGER GMH 3692 Скачать руководство пользователя страница 4

H74.0.14.6C-05 

Operating Manual GMH 3692 

page 4 of 12 

4  Product Description 

4.1  Scope of supply 

The scope of supply includes:  

  GMH 3692 with 9V-battery 

  Operating manual 

The necessary oxygen sensor is chosen separately due to application. 

4.2  Operation and maintenance advice 

1.  Battery operation 

If 

 and 

‘bAt’ are shown in the lower display the battery has been used up and needs to be replaced. 

However, the device will operate correctly for a certain time. If ‘bAt’ is shown in the upper display the 
voltage is too low to operate the device; the battery has been completely used up. 

The battery has to be taken out, when storing device above 50 °C.  
We recommend taking out battery if device is not used for a longer period of time. 
After recommissioning the real-time clock has to be set again.

  

2.  Mains operation with power supply 

When using a power supply please note that operating voltage has to be 10.5 to 12 V DC. 
Do not apply overvoltage!! Cheap 12V-power supplies often have excessive no-load voltage.   
We, therefore, recommend using regulated voltage power supplies.  

Trouble-free operation is guaranteed by our power supply GNG10/3000. 
Prior to connecting the power supply to the mains make sure that the operating voltage stated at the 
power supply is identical to the mains voltage.  

3.  Treat device and sensor carefully. Use only in accordance with above specification. (do not throw, hit 

against etc.). Protect plug and socket from soiling. 

4.3  Start of Operation 

Switch the device on with the key 

“ON OFF”. The Instrument is performing a self diagnosis, during this time 

all display segment are shown. 

Afterwards the instrument signals, if it was user adjusted 

(„

[ORR

“) 

The device starts measurement afterwards.  

 

 

 

Содержание GMH 3692

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Страница 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Страница 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Страница 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Страница 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Страница 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Страница 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Страница 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Страница 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Страница 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Страница 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Страница 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...

Отзывы: