Operating Manual GMH 3692 

page 5 of 12 

4.4  Anschlüsse 

 

1.

 

Output: 

Operation as interface for t he connection of 

galvanically isolated adapters (accessories: GRS 
3100, USB 3100) 

 

Attention: The mode of operation has to be 
configured (p.r.t Chapter 5) and influences the power 
consumption. 

2.

 

Sensor connection 

MiniDIN 

3.

 

Power supply:

 d.c.connector (internal pin Ø 1.9 mm) for external 10.5-12V direct voltage supply 

4.5  Display elements 

 

1 = Main Display:

 

Possible displays: 

  Oxygen concentration in % (% O

Vol) 

  Oxygen  partialpressure (hPa or mmHg) 

(change with 

-key) 

2 = secondary 
display:

 

Display of sensor temperature or absoute 
pressure (alternating, please refer to Chapter 
5.

 L

c

D.2

Special elements:

 

3 = MIN/MAX/HLD:

 

Shows, if  minimum/maximum/ 
memorized measuring value is in display 

4 = ok-arrow:

 

Signals, if oxygen and temperature values 
are stable 

5 =  CAL- arrow:

 

Signals, if automatic calibration is in 
progress 

6 = Logg - arrow  

No function 

7 =  Warning sign:

 

Signals weak battery or other warning 
message 

Die restlichen Pfeile haben in dieser Gerätevariante keine Funktion 

4.6  Pushbuttons 

 

 

On / off key

 

 

Set/Menü 

  Press 2 sec.: (Menu): call configuration menu 
  Press shortly: Change the oxygen display unit  
    (please refer to chapter 5)

 

 

 

min/max when taking measurements: 

press shortly:      min. or max. measured oxygen value and 

referring  temperature and pressure values 
will be displayed 

press for 2 sec.:  the min. or max. value will be deleted 

Configuration

: to enter values, or change settings 

 

Store/Enter 

Measuring:  
with Auto-Hold off: h

old and save current measuring value (‘HLD’ is displayed) 

with Auto-

Hold on: start new measuring, It is finished , when “HLD’ shows in display 

Set/Menu:  

confirm settings, return to measuring 

 

CAL

press shortly: 

display of sensor state rating  

press for 2 sec: 

start sensor calibration 

 

 

Содержание GMH 3692

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Страница 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Страница 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Страница 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Страница 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Страница 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Страница 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Страница 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Страница 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Страница 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Страница 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Страница 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...

Отзывы: