H74.0.14.6C-05 

Operating Manual GMH 3692 

page 12 of 12 

12 Specification 

Measuring ranges 

Oxygen concentration 

0.0 ... 100.0 % O

(Vol)  electrochemical sensors GGO / GOO  

Oxygen partial pressure  0 ... 1100 hPa O

2

 

 

“  

“  

“ 

Sensor temperature 

-5.0 ... + 50.0 °C 

NTC 10k (integr. in GGO / GOO cable) 

Absolute pressure 

10 ... 1200 hPa abs.

 

integrated pressure sensor

 

Accuracy 

(instrument without 
sensor, at 25°C, 1000 
hPa abs)

 

Oxygen concentration 

±0.1 % O

(Vol) 

Oxygen partial pressure  ± 1 hPa 
Sensor temperature 

± 0.1 °C 

Accuracy  

Absolute pressure 

3 hPa or 0.1% of measured value (the higher one to be applied) 

Working conditions 

 

-20 ... 50 °C; 0 ... 95 % r.H. (not condensing) 

Nom. temperature 

 

25°C 

Storage temp. 

 

-20 ... 70 °C 

Connections 

O

2

 & temperature 

6 pole Mini-DIN Socket 

Interface / 
 

serial, (3.5mm audio plug),via isolated adapter GRS3100, GRS3105 or 
USB3100 (accessories) for PC-USB or RS232- connection 

 

external supply 

d.c. connector (diameter of internal pin 1.9 mm) for external 10.5-12V 
direct voltage supply.  (suitable power supply: GNG10/3000) 

Display 

 

4 digit 7-segment 2 lines, additional segments 

Calibration 

automatic 

1 -, 2- or 3-point calibration,  
0%, 100% or ambient air (20.95%) 

GLP 

adjustable calibration intervals (1 to 365 days, CAL warning after 
expiration) 

Alarm 

Buzzer / visual / interface  
2 channels: selectable oxygen unit and temperature 

Additional functions 

Min / max / hold / auto hold 

Housing 

 

Break-proof ABS housing 

 

Protection class 

Front side IP65 

 

Dimensions L*B*H [mm] 
Weight 

Without pressure port: 142 x 71 x 26 mm (L x B x H) 

 

approx. 160 g (incl. battery) 

Power supply 
   Current consumption 

9V-Battery, Type IEC 6F22 (in scope of supply) or external supply 
Ca. 0.6 mA (if Out = Off ca. 0.4mA) 

Change battery indicator 

Automatic at weak battery 

 

and ’ bAt ’ 

Auto-Off-Function 

Device will be automatically switched off if no key is pressed/no 
interface communication takes place for the time of the power-off delay. 
The power-off delay can be set to values between 1and 120 min.; it can 
be completely deactivated. 

EMI 

The device corresponds to the essential protection ratings established in 
the Regulations of the Council for the Approximation of Legislation for 
the member countries regarding electromagnetic compatibility 
(2004/108/EG). Additional fault: <1% 

13 Reshipment and Disposall 

 

Dispense exhausted batteries at destined gathering places.  
This device must not be disposed as “residual waste”. To dispose this device, please send it 
directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly. 

DANGER

  

All devices returned to the manufacturer have to be free of any residual of measuring media and 
other hazardous substances. Measuring residuals at housing or sensor may be a risk for persons 
or environment 

 

Use an adequate transport package for reshipment, especially for fully functional devices. Please 
make sure that the device is protected in the package by enough packing materials. 

 
 

 

Содержание GMH 3692

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Страница 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Страница 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Страница 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Страница 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Страница 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Страница 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Страница 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Страница 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Страница 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Страница 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Страница 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...

Отзывы: