H74.0.05.6C-06 

Operating Manual GMH 3651 

page 12 of 18 

8.2  2 / 3-Point Calibration ('

(AL  2-PT

(AL  3-PT

') 

The sensor will be automatically calibrated to the oxygen content of the atmosphere (20.95%) and one or two 
additional concentrations. 

The calibration can be performed in gases (

[AL A,R

: Air, 100% Oxygen and/ or 0% oxygen like nitrogen) or in 

saturated water solutions (

[AL AQUA

). The water solutions therefore have to be 

“gased” sufficiently with the 

referring gases. Another alternative for 0% is a Sodium sulfite solution ( can also be used at 

[AL A,R

1.  Start calibration: press 

-key for 2 seconds 

2.  First calibration reference: (Pt.1) 

As first reference at a 3-point calibration, the zero reference has to be applied (

NULL

),  

at a 2-point calibration either 100% or 0%(

NULL

). 

The display will show 

, and the referring reference which should be applied: 

-  

NULL

 for 0% oxygen 

-  

0.2

 for pure oxygen or oxygen saturated water

 

As long as the display blinks, no valid reference is recognised by the instrument. 
As soon the values for oxygen and temperature are stable, the calibration of the first point will be finished

The instrument tells you to apply the next reference (possible references are blinking in the display). 

3.  Second calibration reference: (Pt.2) 

The display will show 

, and the referring reference which should be applied: 

-  

A,R

 for ambient air or 

AQUA 

for air saturated water 

-  

0.2

 for pure oxygen or oxygen saturated water 

-  

NULL

 for 0% oxygen 

As long as the display blinks, no valid reference is recognised by the instrument. 
As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the electrode state resulting of the 

successful calibration will be shown for a short time (evaluation in 10% steps:  xx% 

ELE[

). 

At 3-point calibration the instrument tells you to apply the next reference (possible reference is blinking) 

4.  Third calibration reference: (Pt.3) 

The display will show 

, and the referring reference which should be applied: 

As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the sensor state resulting of the successful 

calibration will be shown for a short time (evaluation in 10% steps:  xx% 

ELE[

). 

 

In case of error messages being displayed during the calibration process, please refer to our notes 
at the end of this manual! If a calibration cannot be carried out after an extended period of time, at 
least one of the measuring values is unstable (oxygen partial pressure, temperature).  
Please check your measuring arrangements! 

8.3  Evaluation of Sensor State (

ELE[

Watch sensor state:  press key "CAL" shortly once display show for a short time xx% 

ELE[

It will show the electrode state resulting of the last successful calibration carried out. 
The valuation is displayed in 10 percent steps: 100% means optimal sensor condition. Lower values are 
indicating that the sensor life time will be reached soon.  

Remark:   But also a erroneous pressure may be the cause of low valuation values. 
 

 

Содержание GMH 3651

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Страница 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Страница 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Страница 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Страница 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Страница 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Страница 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Страница 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Страница 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Страница 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Страница 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Страница 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Страница 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Страница 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Страница 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Страница 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Страница 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Страница 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Отзывы: