15 Error and System Messages 

Display 

Meaning 

Remedy 

 

low battery voltage, device will continue to work 
for a short time 

replace battery 

If mains operation: wrong voltage 

replace power supply, if fault continues to 
exist: device damaged 

 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

Check/replace power supply, if fault continues 
to exist: device damaged 

No display 

or 

weird display 

 

Device does not 

react on keys 

low battery voltage 

replace battery 

If mains operation: wrong voltage 

check/replace power supply, if fault continues 
to exist: device damaged 

system error 

disconnect battery or power supply, wait some 
time, re-connect 

device defective 

return to manufacturer for repair 

SENS 

ERRO

 

sensor error: no sensor cable connected 

connect suitable sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.1 

Value exceeding measuring range 

Check: Is the value exceeding the measuring 
range specified? ->value too high! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.2 

Value below display range 

Check: Is the value below the measuring 
range specified? ->value too low! 

Wrong sensor connected 

Check sensor 

Sensor, cable or instrument defect 

return to manufacturer for repair 

ERR.7 

system error 

return to manufacturer for repair 

If 

BAT

“ is flashing, the battery will be exhausted soon. Further measurements are possible for short time. 

If 

BAT

“ is displayed continuously the battery is ultimately exhausted and has to be replaced. Further 

measure

ments aren’t possible any more. 

Messages During Calibration/Adjustment 

>CAL< 

CAL flashing in 

display 

either preset calibration interval has expired or 
last calibration is not valid 

Device has to be calibrated! 

[AL  ERR.1

 

wrong reference point at air  

check sensor and reference gas / solution 

[AL  ERR.2

 

slope too low 

 

reference gas / solution wrong 

check sensor and reference gas / solution 

sensor is defect  

replace sensor / perform maintenance 

[AL  ERR.3

 

slope too high 

 

reference gas / solution wrong 

check sensor and reference gas / solution 

sensor is defect  

replace sensor / perform maintenance 

[AL  ERR.4

 

incorrect calibration temperature 

calibration can only be done at 0…40 °C 

[AL  ERR.5

 

Zero value to low/negative 

 

sensor is defect  

replace sensor / perform maintenance 

[AL  ERR.6

 

zero value to high 

 

reference gas / solution wrong 

check sensor and reference gas / solution 

sensor is defect  

replace sensor / perform maintenance 

[AL  ERR.7

 

incorrect calibration pressure 

check calibration pressure 

[AL  ERR.8

 

signal not stable / timeout 

check sensor and reference gas / solution 

[AL  ERR.9

 

sensor not known: cannot be calibrated

 

check sensor and wiring 

 

 

Содержание GMH 3651

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Страница 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Страница 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Страница 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Страница 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Страница 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Страница 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Страница 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Страница 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Страница 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Страница 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Страница 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Страница 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Страница 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Страница 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Страница 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Страница 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Страница 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Отзывы: