8  Calibration of the Sensor 

In order to compensate for ageing of the sensor, the sensor has to be calibrated at regular intervals.  
The device is equipped with easy-to-use calibration functions: Standard 1-point calibration, sufficient for the 
most applications and special 2 or 3-point calibration, if high accuracy is needed in extreme measuring. 
We recommend to calibrate the sensor at least all 7 days, or to get maximum precision, before each 
measuring series. 
 
If the electrode was dry for one or more days it has to be 'watered' for at least 30 minutes before carrying out 
a new calibration. 
Remove protective flask prior to calibration and wipe diaphragm with a soft pieceof cloth 

8.1  One Point Calibration ('

(AL  1-PT

') 

The 1-point calibration adjusts the sensor to the oxygen content of the atmosphere 
(20.95%). You can choose between three modes of calibration for which the electrode has to 
be prepared accordingly. 

 
Air calibration without accessories ('Cal Air') 

The electrode will be exposed to 

ambient air. 

To protect it 

from draughts, wrap electrode in a paper towel. (before 
calibration, expose sensor at least 15 minutes to the 
ambient air, to let the temperature adjust and to dry 
membrane) 
Depending on the rel. atmospheric humidity [%] and the 
temperature [°C] a small calibration error cannot be 
avoided and will have to be accepted. The cooler the air is, 
the smaller the deviation.  
Recommended temperature:  < 25°C.  
For error compensation please refer to the opposite table. 
Note: A correctly calibrated electrode shows 106 - 108% at air and at room temperature. 

 
Air calibration for highly accurate measurements ('Cal Air') 

The electrode is exposed to 

air with a relative atmospheric humidity of 100%

Proceed as follows: Put some distilled water in a bottle. Close bottle and generate a water steam saturation 
(100% rel. atmospheric humidity) in the remaining air by shaking it vigorously for approx. 3 minutes. Both 
water and air temperatures should be identical. Open bottle and insert electrode so that the diaphragm is in 
the air chamber. 

Attention

: By no means must the diaphragm get wet or be immersed in water. The bottle neck opening 

should only be slightly larger than the electrode diameter; make sure to avoid over pressure in the container. 
Note: A correctly calibrated electrode shows 106 - 108% at air and at room temperature. 

 
Calibration in air saturated water (configuration: 'Cal AqUA') 

This mode of calibration is more difficult than the ones already described. As water is easily oversaturated it 
quickly results in calibration errors. In order to generate air saturated water, pure water (25-30°C) is poured 
into another container from a height of 50 cm. This process is repeated 20 times. In order to let an 
oversaturation evaporate, wait approx. 5 min. before carrying out the calibration (do not forget to stir). 
 

Start calibration: press 

-key for 2 seconds 

The display will show '

A,R

 

, and as soon the values for oxygen and temperature are stable, the 

calibration will be finished. 
Then the electrode state resulting of the successful calibration will be shown for a short time (evaluation in 

10% steps:  xx% 

ELE[

).). 

 

 

 

20% 

40% 

60% 

80% 

100% 

5 °C 

1.007     1.005 

1.003 

1.002 

1.00 

10 °C 

1.01  

1.007 

1.005 

1.002 

1.00 

15 °C 

1.014  

1.01 

1.007 

1.003 

1.00 

20 °C 

1.019   

1.014 

1.009 

1.005 

1.00 

25 °C 

1.026  

1.019 

1.013 

1.006 

1.00 

30 °C 

1.035   

1.026 

1.017 

1.009 

1.00 

35 °C 

1.047     1.035 

1.023 

1.012 

1.00 

40 °C 

1.063     1.047 

1.031 

1.016 

1.00 

Deviations when carrying out an air calibration 
  

without accessories, 

O2-saturation= display value*corr. factor

 

Содержание GMH 3651

Страница 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Страница 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Страница 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Страница 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Страница 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Страница 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Страница 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Страница 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Страница 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Страница 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Страница 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Страница 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Страница 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Страница 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Страница 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Страница 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Страница 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Страница 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Отзывы: