GREISINGER GMH 3610 Скачать руководство пользователя страница 9

H36.0.01.6C-13 

Operating Manual GMH 3610 

page 9 of 12

 

 

10.1  1-Point calibration 

The 1-point calibration adjusts the electrode to the oxygen content of the atmosphere 
(20.95%). Remove protective flask prior to calibration and wipe diaphragm with a soft piece 
of cloth. You can choose between two modes of calibration for which the electrode has to 
be prepared accordingly. We recommend to calibrate for each set of measuring. 
 

Air calibration without accessories  

The electrode will be exposed to 

ambient air. 

To protect it from 

draughts, wrap electrode in a towel or Kleenex. (before calibra-
tion, expose electrode at least 15 minutes to the ambient air, to 
let the temperature adjust and to dry membrane) 
Depending on the rel. atmospheric humidity [%] and the tempera-
ture [°C] a small calibration error cannot be avoided and will have 
to be accepted. The cooler the air the smaller the deviation. Rec-
ommended temperature < 25°C. For error compensation please 
refer to the opposite table. 
Note: A correctly calibrated electrode shows 106 - 108% at air 
and at room temperature. 

 
 
Air calibration for highly accurate measurements  

The electrode is exposed to 

air with a relative atmospheric humidity of 100%

Proceed as follows: Put some distilled water in a bottle. Close bottle and generate a water steam saturation (100% 
rel. atmospheric humidity) in the remaining air by shaking it vigorously for approx. 3 minutes. Both water and air 
temperatures should be identical. Open bottle and insert electrode so that the diaphragm is in the air chamber. 

Attention

: By no means must the diaphragm get wet or be immersed in water. The bottle neck opening should only 

be slightly larger than the electrode diameter; make sure to avoid over pressure in the container. 
Note: A correctly calibrated electrode shows 106 - 108% at air and at room temperature. 
 

Start calibration: press 

-key for 2 sec. 

The display will show 'CAL'; calibration will be automatically completed as soon as the measuring values for oxygen 
and temperature are stable.  
Then the electrode state resulting of the successful calibration will be shown for a short time (evaluation in 10% 
steps). 
 

In case of error messages being displayed during the calibration process, please refer to our notes at the 
end of this manual! If a calibration cannot be carried out after an extended period of time, at least one of 
the measuring values in unstable (oxygen partial pressure, temperature). Please check your measuring 
arrangements! 

 
 

10.2  Valuation of sensor state (ELEC) 

Watch sensor state: 

press key "CAL" shortly once 

display show for a short time xx% ELEC 

It will show the electrode state resulting of the last successful calibration carried out. 
The valuation is displayed in 10 percent steps: 100% means optimal sensor condition. Lower values are indicating 
that the sensor life time will be reached soon.  

Remark:   But also a erroneous pressure entry may be the cause of low valuation values. 

 
 
 

 

 

20% 

40% 

60% 

80% 

100% 

5 °C 

1.007     1.005 

1.003 

1.002 

1.00 

10 °C 

1.01  

1.007 

1.005 

1.002 

1.00 

15 °C 

1.014  

1.01 

1.007 

1.003 

1.00 

20 °C 

1.019   

1.014 

1.009 

1.005 

1.00 

25 °C 

1.026  

1.019 

1.013 

1.006 

1.00 

30 °C 

1.035   

1.026 

1.017 

1.009 

1.00 

35 °C 

1.047     1.035 

1.023 

1.012 

1.00 

40 °C 

1.063     1.047 

1.031 

1.016 

1.00 

Deviations when carrying out an air calibration 
  

without accessories, 

O2-saturation= display value*corr. factor

 

Содержание GMH 3610

Страница 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Страница 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Страница 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Страница 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Страница 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Страница 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Страница 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Страница 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Страница 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Страница 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Страница 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Страница 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Отзывы: