H36.0.01.6C-13 

Operating Manual GMH 3610 

page 6 of 12

 

 

Configuration 

For configuration of the device press 

 -key for 2 seconds: the main menu of the configuration will be called up. 

Use key 

 to select a sub-menu, use the key 

 to actually go into the selected sub-menu and to change parame-

ters. 

Use the keys 

 and 

 to set the individual value for the parameter. Press the key 

 again to memorize the 

changes and to return to the main menu. Use key 

 to leave the configuration. 

 

 

’Pressure ABs.’: Input of Absolute Pressure 

 

500 … 2000 hPa 
abs.

 

The calculated oxygen values will refer to the en-
tered absolute pressure. 

 

 

’Offset’:  Zero Point Displacement of the Temperature

 

 

-3.0 °C ... 3.0 °C 
or 
-5.4 °F ... 5.4 °F 

The zero point of the measurement will be displaced 
for that value set to compensate for sensor and 
measuring device deviations.

 

off: 

Zero point displacement is deactivated (=0.0°)

 

 

’Unit t’:  Selection of Temperature Unit °C /°F

 

 

°C: 

All temperature values in degrees Celsius

 

°F: 

All temperature values in degrees Fahrenheit

 

 

’Power.off’: Selection of Power-Off Delay

 

 

1...120

:

 

Power-off delay in minutes. Device will be automatically 
switched off as soon as this time has elapsed if no key is 
pressed/no interface communication takes place.  
(automatically deactivated for cyclic loggers) 

off: 

automatic power-off function deactivated  
(continuous operation, e.g. in case of mains operation) 

 

’Address’: Selection of Base Address’

 

 

01, 11, 21, …, 91: 

Base address for interface communication. Chan-
nel 1 will be addressed by the set base address, 
channel 2 and 3 will have the following addresses. 
(Example: base address 21 - channel 1 = 21, 
channel 2 = 22, channel 3 = 23)

 

Using the interface converter GRS3105 it is possible to connect several devices to a single interface. As 
a precondition the base addresses of all devices must not be identical. In case several devices are con-
nected via one interface make sure to configure the base addresses accordingly.  
 

 
 
 
 
 
 
 
 
 
 
 
 
 

   

2

Logg

A L

mg/l

ok

mmH g

hPa

SA L

hPa

%O

C A L

max

min

2

5

°C

   

2

Logg

A L

mg/l

ok

mmH g

hPa

SA L

hPa

%O

C A L

max

min

2

5

°C

   

2

Logg

A L

mg/l

ok

mmH g

hPa

SA L

hPa

%O

C A L

max

min

2

5

max

min

2

5

max

min

2

5

Содержание GMH 3610

Страница 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Страница 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Страница 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Страница 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Страница 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Страница 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Страница 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Страница 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Страница 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Страница 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Страница 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Страница 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Отзывы: