H36.0.01.6C-13 

Operating Manual GMH 3610 

page 10 of 12

 

 

11  Error And System Messages  

Display 

Description / Reason 

Remedy 

 

No sensor  
 
or sensor defective 

Connect sensor  
 
sensor defective => return sensor to manufac-
turer for repair 

 

Low battery voltage  
device will continue operation only for 
a short time 

replace battery 

 

Low battery voltage 

replace battery 

- If mains operation: wrong voltage 

replace power supply, if fault continues to exist: 
device damaged 

No display

 

or 

mazy characters

 

- Battery voltage too low 

replace battery 

- If mains op.: power supply defective 

or wrong voltage/polarity 

check/replace mains supply 

- System error 

disconnect battery or power supply, wait for a 
short time, re-connect 

- Device defective 

return to manufacturer for repair 

 

Values exceeding measuring range 
 

Are there any values exceeding the measuring 
range specified? 
 ->measuring value too high 

Sensor/cable defective 

-> replace probe 

 

Values below measuring range 

Are there any values below the measuring 
range specified?  
-> measuring value too low  

Sensor/cable defective 

-> replace probe 

 

System fault 

switch on again: if fault continues to exist, de-
vice is damaged  
-> return to manufacturer for repair 

 

No sensor 

connect suitable sensor 

Error in sensor 

Sensor defective => return to manufacturer for repair 

Temperature display correct,  
oxygen display incorrect 

check: plug in sensor housing connected? 
Open PG-glanding and pull up plug as far as 
possible. 

 

Value could not be calculated 

One measuring variable required for calculation 
is missing (no sensor) or incorrect (over-
flow/underflow) 

 

11.1  Error and System Messages during Oxygen Calibration 

Display 

Description / Reason 

Remedy 

(AL 

ERR. 1 

Wrong temperature 

temperature has to be between 5 and 40°C 

(AL 

ERR. 2 

Wrong atmospheric pressure 

atmospheric pressure has to be between 500 
and 1100 hPa 

(AL 

ERR.3 

Current too low 

membrane dried up / polluted  
=> store electrode in water for 2 hours. 
regenerate or replace electrode. 
check calibration environment (p.r.t. 'How to 
calibrate oxygen electrode.') 

Содержание GMH 3610

Страница 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Страница 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Страница 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Страница 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Страница 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Страница 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Страница 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Страница 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Страница 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Страница 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Страница 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Страница 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Отзывы: