background image

H40.0.01.6C-09 

Operating Manual GMH 3330 

Page 9 of 16

 

 

Measurements Using Combination Measuring Sensor TFS0100  

The TFS0100 has been especially designed to carry out measurements of ambient temperature. All 
TFS0100-probes are interchangeable without recalibration being required. The scope of supply includes 
one sensor to measure relative atmospheric humidity and another one to measure the ambient tempera-
ture T1. 

rel. humidity r.H. [%] 

relative humidity measured in the tip of the probe. Resolution 0.1% 

Ambient temperature T1 

temperature measured in the tip of the probe. Resolution 0.1°C or 0.1°F. 

Other values on display will be calculated by the measuring device (acc. to Mollier diagram): 

Dew point temperature Td

 

Cold air cannot absorb as much steam as warm air. This means that the 

relative 

humidity increases as 

the temperature decreases. If 100% have been reached, the air is saturated with steam; another de-
crease in temperature results in part of the steam condensing to water, becoming visible as fog or precipi-
tation (dew).

 

The dew point temperature indicates at which temperature a 100% saturation would be reached and as of 
when "dew" can be expected. 

Enthalpy h [kJ/kg] 

Enthalpy refers to the energy content of air. This value always refers to dry air at 0° C. I.e. the energy con-
tent of air with a relative humidity of 0% and 0°C is 0kJ/kg. The warmer the air the higher the relative hu-
midity, the higher the energy content. Therefore, more energy is required to heat up humid air than dry air. 

 

All humidity and temperature values calculated from the measuring values refer to a standard 
atmospheric pressure of 1013 mbar. For measuring atmospheric air, the deviations do not 
have to be taken into account. 
When taking measurements in pressure vessels or under similar conditions, the values have to 
be corrected in accordance with a suitable correction table.

  

 

Additional Measurements with NiCr-Ni-Surface Probe at T2: 

 

Surface temperature T2 

The second temperature channel can amongst other things be used to take measurements of surface 
temperatures. 

 

Dew point distance  Td 

This measurement refers to measurements of T1, T2 and relative atmospheric humidity. 
The combination sensor is used to measure the ambient air, whose condition is used to calculate the dew 
point Td. The surface sensor is used to measure surfaces within this ambient air, with  Td stating the 
temperature difference between those measurements and the dew point. 
Example: measuring the ambient temperature results in a Td of 5°C. As long as the surface-temperature 
(T2) of a window exceeds 5°C ( Td > 0°C) the surface won´t sweat! When T2 falls below 5°C, ( Td < 
0°C) it will sweat. 
Other examples for application: detection of 'humid corners', monitoring of heat exchangers, weather fore-
cast etc.. 

 

 

Содержание GMH 3330

Страница 1: ...93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Digital Precision Hand held Measuring Device For Atmospheric Humidity Temperature Dew Point Dew Point Distance Enthalpy and Flow...

Страница 2: ...COMBINATION MEASURING SENSOR TFS0100 9 7 MEASUREMENTS USING FLOW MEASURING PROBES STS005 STS020 10 8 NOTES TO SPECIAL FUNCTIONS 11 8 1 ZERO DISPLACEMENT OFFSET 11 8 2 DISPLAY CORRECTION FACTOR CORR 1...

Страница 3: ...ombined with the combined measuring probe TFS 0100 is designed for room climate meas urements This includes gauging of atmospheric humidity temperature dew point dew point distance and enthalpy Togeth...

Страница 4: ...n lead to malfunctions or destroying of the device and the connected devices This device must not be run with a defective or damaged power supply unit Danger to life due to electrical shock 4 Do not u...

Страница 5: ...g 4 1 Display Depending on the measuring probes sensors connected the following measuring results can be displayed TFS 0100 Main display r H relative atmospheric humidity in Secondary display possible...

Страница 6: ...ing press Menu for 2 sec configuration menu is activated Store Quit Measurement Hold current measuring value HLD in display for flow meas in AVGHold mode start new measurement or handling of logger fu...

Страница 7: ...tc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 Device attached to...

Страница 8: ...ll be multiplied by this factor off Factor is deactivated 1 000 Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time h...

Страница 9: ...ergy con tent of air with a relative humidity of 0 and 0 C is 0kJ kg The warmer the air the higher the relative hu midity the higher the energy content Therefore more energy is required to heat up hum...

Страница 10: ...tions tend to be quite high To be able to display a stable measur ing value two averaging functions have been integrated in the instrument Continuous Averaging The average value displayed has been cal...

Страница 11: ...sses of all devices must not be identical In case several devices will be connected via one interface make sure to configurate the base addresses accordingly Channel 1 will be addressed by the base ad...

Страница 12: ...values a stable value has been de tected and can be taken over by means of the Store key key 6 Then the next calibration step will be displayed 3 Calibration point 2 Put sensor into suitable calibrat...

Страница 13: ...cable defective replace Values below measuring range check are there any values below the measuring range specified meas value too low Sensor cable defective replace System fault switch on again if fa...

Страница 14: ...Windows 7 Windows 8 8 1 Windows 10 Programming examples Visual Basic 4 0 Delphi 1 0 The following interface functions will be supported Channel Dll Code Name function 1 2 3 4 5 6 x x x x x x 0 Read n...

Страница 15: ...lip for table top or sus pended use Weight approx 160 g Interface serial interface 3 5mm jack serial interface can be connected to RS232 or USB inter face of a PC via electrically isolated interface a...

Страница 16: ...dequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Bat...

Отзывы: