background image

H40.0.01.6C-09 

Operating Manual GMH 3330 

Page 13 of 16

 

 

10.2  System and Error Messages  

System or error 
messages 

Description / Reason 

Remedy 

 

no probe/sensor connected 

connect probe/sensor 

probe/sensor damaged 

probe/sensor defective 

 return to manufacturer for 

repair 

 

Low battery voltage, device will 
only continue operation for a short 
time 

replace battery 

 

Low battery voltage 

replace battery 

If mains operation: wrong voltage 

replace power supply, if fault continues to exist: device 
damaged 

No display 

or 

mazy characters

 

Battery voltage too low 

replace battery 

If mains op.: power supply defec-
tive or wrong voltage/polarity 

check/replace power supply 

System error 

disconnect battery or power supply, wait for a short time, 
re-connect 

device defective 

return to manufacturer for repair 

 

Values exceeding measuring  
range 

Check: are there any values exceeding the measuring 
range specified? ->meas. value too high 

Sensor/cable defective 

-> replace 

 

Values below measuring range 

check: are there any values below the measuring 
range specified? ->meas. value too low 

Sensor/cable defective 

-> replace 

 

System fault 

switch on again: if fault continues to exist, device is 
damaged 
-> return to manufacturer for repair 

Instrument not within working tem-
perature 

keep working temperature in between -25...50°C 

 

No probe/sensor existing  
or probe/sensor defective 

connect probe/sensor; 
probe/sensor damaged -> return to manufacturer for 
repair 

 

Value cannot be calculated 

One measuring variable required for calculation is miss-
ing (no sensor) or incorrect (overflow/underflow) 

 

10.3  System and Error Messages during TFS0100 Calibration  

Error or system 
messages 

Description / Reason 

Description / Reason 

 

Deviation to high (zero point) 

correct humidity variable? 
no -> probe no longer within permissible tolerances, 
return to manufacturer for recalibration. 

 

Difference point1-point2 too small 

difference has to be at least 40% if values are entered 
manually select suitable  

 

Incorrect temperature 

calibration is only permissible in the temp. range from 
5 ... 40°C 

m b a r

m b a r

m b a r

m b a r

Содержание GMH 3330

Страница 1: ...93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Digital Precision Hand held Measuring Device For Atmospheric Humidity Temperature Dew Point Dew Point Distance Enthalpy and Flow...

Страница 2: ...COMBINATION MEASURING SENSOR TFS0100 9 7 MEASUREMENTS USING FLOW MEASURING PROBES STS005 STS020 10 8 NOTES TO SPECIAL FUNCTIONS 11 8 1 ZERO DISPLACEMENT OFFSET 11 8 2 DISPLAY CORRECTION FACTOR CORR 1...

Страница 3: ...ombined with the combined measuring probe TFS 0100 is designed for room climate meas urements This includes gauging of atmospheric humidity temperature dew point dew point distance and enthalpy Togeth...

Страница 4: ...n lead to malfunctions or destroying of the device and the connected devices This device must not be run with a defective or damaged power supply unit Danger to life due to electrical shock 4 Do not u...

Страница 5: ...g 4 1 Display Depending on the measuring probes sensors connected the following measuring results can be displayed TFS 0100 Main display r H relative atmospheric humidity in Secondary display possible...

Страница 6: ...ing press Menu for 2 sec configuration menu is activated Store Quit Measurement Hold current measuring value HLD in display for flow meas in AVGHold mode start new measurement or handling of logger fu...

Страница 7: ...tc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 Device attached to...

Страница 8: ...ll be multiplied by this factor off Factor is deactivated 1 000 Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time h...

Страница 9: ...ergy con tent of air with a relative humidity of 0 and 0 C is 0kJ kg The warmer the air the higher the relative hu midity the higher the energy content Therefore more energy is required to heat up hum...

Страница 10: ...tions tend to be quite high To be able to display a stable measur ing value two averaging functions have been integrated in the instrument Continuous Averaging The average value displayed has been cal...

Страница 11: ...sses of all devices must not be identical In case several devices will be connected via one interface make sure to configurate the base addresses accordingly Channel 1 will be addressed by the base ad...

Страница 12: ...values a stable value has been de tected and can be taken over by means of the Store key key 6 Then the next calibration step will be displayed 3 Calibration point 2 Put sensor into suitable calibrat...

Страница 13: ...cable defective replace Values below measuring range check are there any values below the measuring range specified meas value too low Sensor cable defective replace System fault switch on again if fa...

Страница 14: ...Windows 7 Windows 8 8 1 Windows 10 Programming examples Visual Basic 4 0 Delphi 1 0 The following interface functions will be supported Channel Dll Code Name function 1 2 3 4 5 6 x x x x x x 0 Read n...

Страница 15: ...lip for table top or sus pended use Weight approx 160 g Interface serial interface 3 5mm jack serial interface can be connected to RS232 or USB inter face of a PC via electrically isolated interface a...

Страница 16: ...dequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Bat...

Отзывы: