22
Using the IPM-400 as a Source
(cont’d)
Calibrate Frequency Inputs
Display:
kHz, Hz, CPM
This function provides pulses into frequency measuring instruments. The meter output is a
zero crossing square wave from –1 V to 5 V amplitude. Available ranges are from 0.01 kHz to
15.00 kHz, 1 Hz to 1500 Hz and from 1 CPM to 1500 counts per minute (CPM). CPM is used
to simulate extremely slow frequency signals with greater resolution. For example, 10 Hz is
equivalent to 600 CPM. To convert from CPM to Hz, divide by 60. To convert from Hz to CPM,
multiply by 60.
1.
Turn the Selector knob to
FREQ
.
2.
Connect the black and red leads to the meter as shown.
3.
Repeatedly press the
DISPLAY/SOURCE/READ
button until “SOURCE” appears on the
display.
4.
Press the
TYPE/ENG UNITS
button to select the desired function.
5.
Disconnect one or both input wires from the device being calibrated.
6.
Connect the red lead of the meter to the plus (+) input of the device and the black lead to
the minus (–).
7.
With the toggle switch at
SET
, adjust the output to the desired value.
8.
Flip the toggle switch to
HI
or
LO
to use the stored values. See the Storing Output Values
section to change the stored values.
Frequency Receiver Input
Flow Meter
Controller
Transmitter
Computer
Logger
DCS
Содержание IPM-400
Страница 46: ...46 Greenlee Textron Subsidiary of Textron Inc 4455 Boeing Dr Rockford IL 61109 2988 815 397 7070...
Страница 92: ...92 Greenlee Textron Subsidiary of Textron Inc 4455 Boeing Dr Rockford IL 61109 2988 815 397 7070...
Страница 138: ...138 Greenlee Textron Subsidiary of Textron Inc 4455 Boeing Dr Rockford IL 61109 2988 815 397 7070...
Страница 139: ...IPM 400 139 Greenlee Textron Subsidiary of Textron Inc 4455 Boeing Dr Rockford IL 61109 2988 815 397 7070...