Common Specifications of the terminal in the input section
* The following specifications are common to GL840-M and GL840-WV.
Description
Item
Measurement
accuracy
20/50/100/200/500 mV, 1/2/5/10/20/50/100 V, 1-5 V F.S.
Thermocouple : K, J, E, T, R, S, B, N, C (WRe5-26)
Resistance bulb : Pt100, JPt100, Pt1000 (IEC751)
Measurement range : 100°C, 500°C, 2000°C
0 to 100% (voltage 0 V to 1 V scaling conversion)
Voltage
Temperature
Humidity
At least 90 dB (50/60 Hz; signal source 300Ω or less)
At least 48 dB (with +/- terminals shorted)
Gain : 0.01% of F.S./°C
* Occurs when sampling speed is 10 ms, 20 ms or 50 ms.
Zero : 0.02% of F.S./°C
Temperature coefficient
16-bit Delta-Sigma A/D converter (Effective resolution: approx. 1/40,000 of ± range)
A/D converter
Common mode rejection ratio
Noise
Number of input channels M3 screw type, 20 channels (maximum 200 channels with extension unit)
Method
Photo MOS relay scanning system, all channels isolated, balanced input
Specifications of input section
(GL840-M with standard terminal)
Description
Item
•
Voltage : ±0.1% of F.S.
Measurement accuracy *
1
(23°C ±5°C)
• When 30 minutes or
more have elapsed after
power was switched on
• Sampling 1 s/20 ch
• Filter ON (10)
• GND connected
Between input terminal/GND : 60Vp-p
Between input terminal/input terminal : 350Vp-p/1 minute
Between input terminal/GND : 350Vp-p/1 minute
Between input terminal/input terminal : 60Vp-p
B/– terminals : 20mV to 2V range (60Vp-p)
5V to 100V range (110Vp-p)
Maximum input voltage
Withstand voltage
*1: Thermocouple diameters T, K: 0.32 Φ, others: 0.65 Φ
* 3-wire system
•
Resistance bulb
Type
Measurement Accuracy
Measurement
Temperature Rang
R/S
B
K
E
T
N
C (W)
Reference contact compensation
accuracy
J
0≤TS≤100˚C
±5.2˚C
±3.0˚C
± (0.05% of rdg +2.0˚C)
± (0.05% of rdg +2.0˚C)
± (0.05% of rdg +2.0˚C)
± (0.05% of rdg +2.0˚C)
± (0.05% of rdg +1.0˚C)
± (0.05% of rdg +2.0˚C)
± (0.05% of rdg +1.0˚C)
± (0.1% of rdg +1.5˚C)
± (0.1% of rdg +0.5˚C)
± (0.05% of rdg +1.0˚C)
± (0.1% of rdg +2.0˚C)
± (0.1% of rdg +1.0˚C)
± (0.1% of rdg +1.5˚C)
±3.5˚C
±2.7˚C
±1.7˚C
±0.5˚C
100<TS≤300˚C
R: 300<TS≤1600˚C
S: 300<TS≤1760˚C
400≤TS≤600˚C
600<TS≤1820˚C
-200≤TS≤-100˚C
-100<TS≤1370˚C
-200≤TS≤-100˚C
-100<TS≤800˚C
-200≤TS≤-100˚C
-100<TS≤400˚C
-200≤TS≤-100˚C
-100<TS≤100˚C
100<TS≤1100˚C
-200≤TS<0˚C
0≤TS≤1300˚C
0≤TS≤2000˚C
•
Thermocouple
Type
Applied current
Accuracy
Measurement
Temperature Range
Pt100
JPt100
Pt1000
-200 to 850°C
±1.0°C
1mA
-200 to 500°C
±0.8°C
1mA
-200 to 500°C
±0.8°C
0.3mA
21