Heat and Massage / Heat Only Unit Troubleshooting Guide

The cables and connections need to be validated. This will require the services of a maintenance 
person and a clear working area.
1.  INSPECTION: Perform the following checks. If a working recliner of the same model is available, 

use it for comparison.
a)  Is the AC cord properly connected to the power supply? Is the cable from the power supply 

to the control module junction box in place?

b)  With the power supply AC cord plugged into a live AC outlet, does the small light to the 

left of the white seat symbol (located on the switch) light up when the switch is pressed 
repeatedly or is there any indication of massage operation?

  WARNING: Disconnect the power cord from the AC outlet before performing the following 

steps!

c)  Is there any sign of external damage to the handheld pendant or wire?
d)  Is there any sign of external damage to the recliner?
e)  Do any of the cables show any sign of damage or pinch points?
f)  Are all cable connections to the control module in place? Are the connectors fully engaged?
g)  Are any of the connectors damaged?
h)  Are the wires within the connectors attached to the pins? Each wire may have to be tugged 

lightly to check if the wire is still crimped to the pin.

2. POST-INSPECTION:

a)  If improper or loose connections are found, reconnect them firmly (unless a loose wire has 

been found). Plug the power supply AC cord into a live wall socket and operate the unit 
through all functions.

b)  If the handheld pendant or wire is damaged, notify Graham-Field so that a handheld 

pendant can be sent for replacement.

c)  If damage to any other electrical cable or component, or disconnected wires are found, 

please notify Graham-Field.

3. ACTIONS:

a)  If full operation is restored, please advise Graham-Field of your successful outcome.
b)  If the fault cannot be corrected by simple re-connection, please inform Graham-Field of the 

results of your inspection in detail so that we may determine the appropriate action to be 
taken.

If you have any questions about the setup and / or operation of the Heat and Massage / Heat Only 
Unit, call the Graham-Field Technical Support Group at 1.770.368.4700.

Содержание LUMEX FR566G Series

Страница 1: ...cliner with Heat and Massage Series FR566DGH Deluxe Clinical Care Recliner with Drop Arm with Heat and Massage Series FR566GHO Deluxe Clinical Care Recliner with Heat Only Series FR566DGHO Deluxe Clin...

Страница 2: ...OWN AS SHOCK POSITION 11 DROP ARM FUNCTION SERIES FR566DG ONLY 12 TO OPERATE THE DROP ARM FEATURE 12 RECLINERS WITH HEAT AND MASSAGE OR HEAT ONLY FEATURE 13 INSTALLATION AND OPERATING GUIDE 13 DESCRIP...

Страница 3: ...must pay attention to and abide by There are additional safety statements in other chapters or sections which may be the same as or similar to the following or specific to the operations Please note t...

Страница 4: ...cept during transport WARNING Unauthorized modification of your recliner or the use of non Lumex replacement parts may change the structure of the recliner and could create a hazardous condition which...

Страница 5: ...side panel Trendelenburg release bar padded push handle total locking caster total of 3 on recliner black locking tab rear directional swivel locking caster right rear leg only green locking tab Clini...

Страница 6: ...mechanism 2 Remove two pre installed threaded knobs from back one on each side 3 Carefully guide the upholstered back onto the mechanism inserting both bayonet links at the same time through the two c...

Страница 7: ...fied by black actuator tabs that lock both the wheel and the swivel To activate the total lock simply press down on the BLACK actuator tab until it locks To release the total lock press on the raised...

Страница 8: ...le disengaged position Side Table SIDE TABLE OPERATION ALL MODELS To Raise Table Raise the side table all the way up and slide in toward the armrest until tabs engage slots in mount bracket Press down...

Страница 9: ...the armrest Make sure side panel is properly aligned and that the top edge is behind the locking tabs Press side panel into frame sides Make sure that both push buttons are protruding into side panel...

Страница 10: ...nd reading To Achieve Second Position First Recline Position Reclining by patient place hands on armrests and push body against recliner back to recline yourself to desired position To bring yourself...

Страница 11: ...ndle with the other hand Push back on push handle to recline To return grasp push handle and pull forward to raise the recliner back Trendelenburg Position Also Known as Shock Position Info This posit...

Страница 12: ...are locked are at the same height in order to reduce the amount of lifting required to transfer the patient and are as close to one another as possible Failure to heed these warnings could result in p...

Страница 13: ...se of this product Description The Lumex Healthcare Seating Heat and Massage Heat Only Unit is a low voltage 12V Heat and Massage Heat Only Unit controllable by the user and or caregiver via a handhel...

Страница 14: ...Only Installation Instructions 1 Place the back cushion over the vertical bayonet lugs of the recliner mechanism Secure the back on each side with the threaded screw knobs per the previous Upholstere...

Страница 15: ...g tabs snap into place as shown above Connector Security Lock 5 If you need to replace the Hand Pendant remove the small white connector security lock as shown above 6 Remove the tape holding the back...

Страница 16: ...0200RevA20 SEE OPERATING MANUAL FOR COMPLETE INSTRUCTIONS HIGH HEAT PRESS 1 TIME MEDIUM HEAT PRESS 2 TIMES LOW HEAT PRESS 3 TIMES 3 The Heat and Massage Pendant Selector Switch Faceplate shown at abov...

Страница 17: ...The Heat Function Settings are as follows a Press selector switch for Setting 1 High Heat 104 F 40 C brightest LED b Press selector switch AGAIN for Setting 2 Medium Heat 101 F 38 6 C bright LED c Pr...

Страница 18: ...ts f Are all cable connections to the control module in place Are the connectors fully engaged g Are any of the connectors damaged h Are the wires within the connectors attached to the pins Each wire...

Страница 19: ...e 1 part isopropyl alcohol to 1 part water Info Using alcohol on a regular basis can affect the properties of the product Once a month you can use a soft bristle brush for a deeper cleaning Info Light...

Страница 20: ...ning and maintaining noncritical medical equipment In the absence of a manufacturer s cleaning instructions follow certain procedures Clean noncritical medical equipment surfaces with a detergent disi...

Страница 21: ...ired by facility policy or after cleaning up large spills of blood or other body substances Clean mops and cloths after use and allow to dry before reuse or use single use disposable mop heads and clo...

Страница 22: ...ent waste groups such as metal cable electronic recoverable resource and plastic for recycling or combustion Most plastic components are provided with a plastic types code and fiber content to aid sor...

Страница 23: ...d on Lumex Recliner product labels MANUFACTURER FRAGILE HANDLE WITH CARE CE MARK CAUTION EC REP EUROPEAN AUTHORIZED REPRESENTATIVE UDI UNIQUE DEVICE IDENTIFIER GENERAL WARNING SIGN KEEP DRY ELECTRICAL...

Страница 24: ...t from Seat 8 20 3 cm 8 20 3 cm Seat Height 22 55 9 cm 22 55 9 cm Seat Width 20 1 2 52 1 cm 20 1 2 52 1 cm Seat Depth 20 50 8 cm 19 48 3 cm Depth Fully Reclined in Trendelenburg 75 190 5 cm 76 193 0 c...

Страница 25: ...case the warranty shall expire on the earlier of warranty period or the expiration date OBTAINING WARRANTY SERVICE A GF Customer Service Representative must authorize warranty service Please contact...

Страница 26: ...INCOME LOSS OF USE DOWNTIME COVER OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the s...

Страница 27: ...ide 18 I Important safety precautions 3 Info statement significance 3 Intended Use 3 Introduction 3 L Limited warranty 25 N NOTICE statement significance 3 O Operation 7 P Product description 5 R Recl...

Страница 28: ...field com 2012 GF Health Products Inc All Rights Reserved Lumex is a trademark of GF Health Products Inc Tente is a trademark of TENTE GmbH Co FED REP GERMANY GF Health Products Inc is an ISO 13485 20...

Отзывы: