SPECIFICATIONS
PHYSICAL AND ELECTRICAL
g
Workstation: 146cm x 53cm x 69cm, 130kg; aluminum frame, vertical mounting side rails on both sides, nylon internal tubing
g
Top shelf: 54 cm (W) x 37 cm (D), maximum load 35 kg with even load distribution; Epoxy powder coated pressed aluminum;
monitor brackets bolt items to the top of the top shelf
g
Drawer: 10cm x 29cm x 32cm in A, 11cm x 29cm x 24cm in B; 35kg weight limit, stainless steel, removable for cleaning.
g
Work surface: Stainless Steel, 46 cm (W) x 31 cm (D) x 88 cm (H) above ground (removable) in A, 46 cm (W) x 26 cm
(D) x 88 cm (H) in B
g
Mains power supply: 220V, 50-60Hz
g
Power: 500 Watts/220v = 2.27 amps
g
Oxygen analyzer power: CE-marked, fused and medical grade. Mains powered with battery backup for ten hours of use.
g
Automatic over/under voltage mains power isolator protects oxygen concentrator, and sockets for monitor and accessories
g
3.5” TFT touch screen for oxygen percentage display and alarm setting
g
Membrane switch for oxygen analyzer power
g
Casters: 150 mm diameter single wheel antistatic casters, front lockable casters
OXYGEN SUPPLY AND MONITORING
g
Oxygen concentrator flow rate: 0.1 to 10.0 liters per minute
g
Maximum oxygen concentration: 95%
NOTE:
The oxygen concentration may vary according to ambient humidity and maintenance of the air filter
g
Alternative sources: cylinder yoke, pipeline, or other external flow regulated portable oxygen source
g
Automatic room air intake when patient minute volume exceeds supply gas flow
g
Accuracy of glass rotameter for O
2
: +/- 2.5% when using 100% O
2
g
Integrated inspiratory oxygen analysis uses a MOX-3 oxygen sensor
g
Calibration for room air (21%) and 100% oxygen
g
Pressure sensitive apnea or high flow alarm
g
Adjustable minimum and maximum oxygen alarm settings
g
Up to ten hour analyzer battery backup, trickle recharge from mains power
g
Membrane keypad and touch screen
OPTIONAL USE OF NITROUS OXIDE
g
Sources: pipeline or cylinder
g
Hypoxic cut-off: A solenoid automatically shuts off nitrous oxide delivery if supply gas O
2
level falls below a minimum of 25%
g
Flow of N
2
O stops if electrical power fails
g
Sight and touch differentiated flow control dial per ISO standards
g
Accuracy of glass rotameter for N
2
O: +/- 2.5%
VENTILATION
g
Manual bellows for adult and pediatric use made from durable, long-lasting silicone rubber
g
1600cc capacity
g
Inspiratory pressure relief of 55 cmH
2
O for pressure created during mechanical ventilation
g
Aluminum bellows block
g
Silicone balloon inflating valve at the expiratory port
g
15mm female/22mm male ISO standard taper connection for breathing circuit, 30mm male ISO standard taper connection for AGSS
scavenging equipment
VAPORIZERS
g
Stainless steel and plated brass construction
g
Draw over flow type
g
Pour filler type.
g
Separate units for isoflurane, sevoflurane, and halothane per ISO standards
g
Agent delivery range: 0.5% to 4%
g
120ml capacity
4
Содержание Universal Anaesthesia Machine
Страница 1: ...Universal Anaesthesia Machine Maintenance Manual...
Страница 14: ...12...
Страница 38: ...36 13 16 28 17 18 24 19 20A 20B 19A 30 47 10 11 12 14 16 13 24 1 44 3B 2A 26 7 21 9 8 31 42 43 3A 2B 20 B...
Страница 39: ...37 2 3 6 41 35 33 32 36 32 34 38 37 35 33 32 36 41 32 6 4 5 46 15 45 49 48 B B...
Страница 40: ...38 41 36 37 39 6 40...
Страница 41: ...39 25 27 29...
Страница 48: ...NOTES...
Страница 49: ......