6.
Attach patient circuit to inspiratory and expiratory ports.
7.
Attach green plastic 30M-30F connector to the scavenging hose by inserting the male side into the hose. The female
side attaches to the one-way expiratory port underneath the Fenton balloon.
8.
Connect 30mm hose to scavenging port, or attach standard scavenging device.
9.
Operate the bellows, check that it fills freely with air from the air inlet and observe the Fenton balloon moves freely in it’s
housing when the bellows is moved. Occlude the Y-piece and check airway pressure reaches 45-55 cmH
2
O.
10.
Connect and turn on mains power by pressing the green switch on the back of the UAM. Wait about 10 seconds for the
internal supply to come on, and then check the screen to see if the oxygen analyzer shows green. This indicates the
mains supply is working correctly.
NOTE:
GRADIAN HEALTH SYSTEMS STRONGLY RECOMMENDS CONNECTING THE UAM TO A 2000-WATT
VOLTAGE STABILIZER.