31
INSPECTING AND CLEANING THE PRESSURE RELIEF VALVE:
5.
Inspect all of the parts to ensure that there are no stains, debris, moisture or rust. Inspect the condition of the
O-ring and make sure it is not cracked or deformed.
6.
To clean, place all the parts in a small bowl with isopropyl alcohol and let them soak for a few minutes.