Setup
Setup
Setup Screen
Screen
Screen 4
4
4
Use this screen to calibrate your meter and
to view or set your meter k-factor. See
, for procedure.
Figure 9 Setup Screen 4
Key
Key
Key
Enter the screen to set or change
preferences.
Press to activate a field for editing or
to accept the highlighted selection on a
dropdown menu.
Move to the right when editing number
fields. Press again to accept the entry
when all digits are correct.
Start the calibration.
Stop the calibration.
Displays the volume measured by the
system for the calibration test.
Enter the actual volume in the cylinder
from the calibration test.
Displays the meter k-factor. User can
set the k-factor manually. The system
automatically updates to the correct
k-factor when the meter is calibrated.
Exit data editing.
Move between Setup Screens, fields on
a screen, or to increment/decrement the
digits when editing number fields.
Setup
Setup
Setup Screen
Screen
Screen 5
5
5
Use this screen to set your modbus preferences
for ports 1 and 2. Note that ports 5 and 6 are used
as modbus master devices for connecting to other
Informer (or ProCrontol 1KE) modules.
Figure 10 Setup Screen 5
Key
Key
Key
Modbus mode. Select off or Slave from
the dropdown options.
Turn off Modbus functionality if
not used.
Use Informer as Modbus slave
device.
Enter or change the Modbus address.
Value is between 1 and 247.
Select serial port baudrate from the
dropdown options: 9600, 19200, 38400,
57600, or 115200.
Select serial port parity from the
dropdown options: NONE, ODD, or
EVEN.,
Exit data editing.
Move between Setup Screens, fields on
a screen, or to increment/decrement the
digits when editing number fields.
3A2040H
27