Immunity Test
IEC 60601-1-2
Test Level
Compliance level
Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst. IEC
61000-4-4
2kV for power supply
lines
2 kV for power supply
lines N/A
Main power quality should be that
of a typical commercial and/or
hospital environment.
Surge
IEC 61000-4-5
1kV line-to-line
1kV line-to-line
Main power quality should be that
of typical commercial and/or
hospital environment.
Voltage
dips,
short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
0% U for 0,5 cycle 70% U
For 25/30 cycles
0% U for 0,5 cycle 70% U
For 25/30 cycles
Main power quality should be that
of a typical commercial and/or
hospital environment. If the user of
the Umana T1 requires continued
operation during power main
interruptions, it is recommended
that
the
Umana
T1
image
intensifier is powered from an
uninterruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
If image distortion occurs, it may
be necessary to position the
Umana T1 image intensifier
further from sources of power
frequency magnetic fields or to
install magnetic shielding. The
power frequency magnetic field
should be measured in the
intended installation location to
ensure that it is sufficiently low.
NOTE: U is the A/c main’s voltage prior to application of the test level.
The Umana T1 is suitable for use in a specified electromagnetic environment. Users should ensure that it is used in an
electromagnetic environment as described below:
Immunity Test
IEC 60601-1-2
Test Level
Compliance level
Electromagnetic Environment
Guidance