umanamedical.com
Page 23 of 27
Use-by Date
Enclosures of the Umana T1 Heart Monitor are protected against the ingress of solid foreign
objects of 1mm diameter and greater and are protected against water projected by a nozzle
(diameter 6.3mm) at a distance of 2.5-3 m.
Class II Medical Equipment
Type CF Applied Part
15. EMC Information
The Umana T1 heart Monitor and the T1 Tempo Dongle are suitable for use in a specified electromagnetic environment. Users
should ensure that it is used in an electromagnetic environment as described below:
Emissions test
Compliance
Electromagnetic Environment Guidance
RF Emissions CISPR 11
Group 1
The Umana T1 only uses RF energy for its
internal
function.
Therefore,
its
RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF Emissions CISPR 11
Class B
The Umana T1 is suitable for use in all
establishments,
including
domestic
establishments
and
those
directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic Emissions IEC 61000-3-2
Class A
The Umana T1 is suitable for use in a specified electromagnetic environment. Users should ensure that it is used in an
electromagnetic environment as described below: