8.1. Preparing the T1 Heart Monitor
•
Charge the T1 Heart Monitor for 4 hours using the provided charger and charging adapter only. Connect the charger to the
adapter, then connect the adapter to the middle magnetic connector of the T1 Heart Monitor. Ensure corrected orientation
accordingly to the plastic notch in the adapter. Ensure that the battery charger is switched ON from the mains supply.
•
To turn T1 Heart Monitor ON, press the power button located on back of device.
•
To turn T1 Heart Monitor OFF, press and hold for 3 seconds the power button located on back of device.
To conserve battery life, turn T1 Heart Monitor ON only at the point of connecting to T1 App. Keep T1 Heart Monitor ON
for the whole duration of the Holter or Monitoring Sessions. The T1 Heart Monitor becomes inactive while charging, as
a safety feature.
8.2. Preparing Patient for the SmarterSkin or Smar Sensor Placement
•
Maintain good infection control practice by washing hands prior to patient contact.
•
Patients with chest hair should have hair at placement site removed, ideally not less than 4 hours before.
Tear open the sealed silver pouch containing the sensor, protective film and wipe. Clean placement site with the alcohol wipe
provided and leave to dry.
Apply SmarterSkin
TM
Sensor on intact clean dry skin only. Abrasive
pastes specifically manufactured for ECG tests may be used to rub on
the skin to increase sensor adherence and remove unwanted dead skin
cells. Such pastes help reduce skin impedance for improved quality of
recordings. Ensure product compatibility with patient prior to use. Ensure
no abrasions are caused. If the skin is oily or the patient applied any
creams or lotions, use a wipe or abrasive paste to clean placement site.
Ensure product compatibility with patient prior to use.
8.3. Application of SmarterSkin or Smar Sensor
•
Sensors are provided in various packs i.e., Getting Started packs (SmarterSkin or Smar Sensors) and Sensor Packs
(SmarterSkin or Smar sensors)
•
Tear open the sealed si
lver pouch containing the SmarterSkin™ / Smar™ Sensor and Protective Film.
•
Remove backing liners from SmarterSkin™ / Smar™ Sensor, following their arrow instructions to expose adhesive side.
•
Place adhesive side of SmarterSkin™ / Smar™ Sensor onto patient’s left side in the third intercostal space
perpendicular to the midclavicular line.
•
Ensure that SmarterSkin™ / Smar™ Sensor is placed with the 3 snaps creating a V shape, and the middle electrode
as vertically straight as possible.
•
Ensure no air bubbles are trapped between SmarterSkin™ / Smar™ Sensor and skin.
•
Remove backing liners from Protective Film to expose adhesive side.
•
Place adhesive side of Protective Film over SmarterSkin™ / Smar™ Sensor.
•
Ensure punched
holes lie directly over the 3 snaps found on the SmarterSkin™ Sensor, leaving them exposed. Also
ensure that the alignment is correct such that the other holes are not covered to avoid issues during use of the T1
monitor and its accessories.