2.  The  far  left  digit  will  be  blinking.  The  DISPLAY  button 

can then be pressed to select the digit location and the 

CALIBRATE button can be pressed to scroll the desired 

value at the blinking position. Edit the K-factor shown to 

the desired value. Acceptable K-factor range is 0000.1 to 

9999.9.

3.  After the new value has been entered, momentarily press 

and release both buttons. “CALEND” will be momentarily 

displayed. Unit is now ready for use.

4.  Notice that the upper display line, the “FAC” icon and all 

the units of measure have disappeared.

Alternate units of measure are not selectable when the meter 

is operating with field calibration. This calibration is a unique 

single-point calibration for the meter and/or application.
NOTE: To return to factory calibration (FAC), press and hold both 

CALIBRATE and DISPLAY buttons for about 3 seconds, 

until  FAcCAL  is  displayed.  Then  release  buttons.  Unit 

should return to normal operation and FAC icon is visible.

NOTE: If the field calibration mode is entered and NO fluid is 

dispensed, then upon leaving, the computer will use data 

from the last successful field calibration.

T A B L E   1

 

 

 

Typical K-Factor

  Model 

Size 

(pulses / gallon)

 

G2_05 

1/2 in. 

2500

 

G2_07 

3/4 in. 

1100

 

G2_10 

1 in. 

560

 

G2_15 

1-1/2 in. 

215

 

G2_20 

2 in. 

100

 

G2P05 

1/2 in. 

2400

 

G2P10 

1 in. 

540

  G_T-051 

1/2 in. 

10000

  G_T-075 

3/4 in. 

3750

  G_T-100 

1 in. 

896

  G_T-150 

1-1/2 in. 

350

  G_T-200 

2 in. 

181

  G_T-300 

3 in. 

50

  G_P-050 

1/2 in. 

10000

  G_P-051 

1/2 in. 

10000

  G_P-075 

3/4 in. 

3750

  G_P-100 

1 in. 

896

  G_P-150 

1-1/2 in. 

340

  G_P-200 

2 in. 

181

  GM001 

1/8 in. 

5855

  GM002 

1/4 in. 

3785

  GM003 

1/4 in. 

1514

  GM005 

1/2 in. 

424

  GM006 

3/4 in. 

197

  GM007 

1 in. 

197

  GM010 

1 in. 

136

  GM015 

1-1/2 in. 

55

  GM020 

2 in. 

25

Field Calibration Procedures

(Correction Factor Method)

If  the  display  was  purchased  with  a  meter  body  then  the 

field calibration method set at the factory is the Correction 

Factor method.
This method allows the user to tweak the factory calibration 

by a percent that represents application, fluid or plumbing 

differences.
1.  To calibrate, press and hold the CALIBRATE and DISPLAY 

buttons for about 3 seconds until you see FLdCAL. Release 

both buttons and you will see CF-00.0. You are now in the 

field calibration mode and values from -99.9% to +99.9% 

can be entered.

2.  The +/– position appears either as an “underscore” char-

acter for plus, or as a “hyphen” character for minus. The 

DISPLAY button selects the position and the CALIBRATE 

button toggles this character.

3.  The DISPLAY button can then be pushed to select the 

numeric positions. Press the CALIBRATE button to scroll 

from 0 to 9. Enter the percentage of change you want the 

display to correct. When satisfied with the value, press 

both CALIBRATE and DISPLAY buttons simultaneously.  

CALEnd will be displayed and unit will go back to normal 

operation, less the FAC (factory calibration) icon.

4.  All enabled units-of-measure remain visible and select-

able – the entered correction will be applied to all enabled 

units.

5.  To return to factory calibration (FAC), press and hold both 

CALIBRATE and DISPLAY buttons for about 3 seconds 

until  FAcCAL  is  displayed.  Then  release  buttons.  Unit 

should return to normal operation and FAC icon is visible.

Setting 4-20 mA Endpoints

The  4-20  mA  endpoint  settings  are  independent  from  the 

display calibration. If you reset the response time you MUST 

reset the 4-20 mA endpoints.
All units are shipped with the following items preset:
  •  4 mA setpoint = 10 Hz
  •  20 mA setpoint = 1000 Hz
  •  Response time = 0.7 seconds
Any  new  values  you  set  for  these  items  are  automatically 

saved when the transmitter is powered down, and automati-

cally restored the next time power is applied.

Procedure

Before you start, the fluid pumping system should be ready 

to make two simple calibrating runs, first at the lowest an-

ticipated flowrate, and then the second at the highest antici-

pated flowrate. Position yourself so you can easily operate 

the transmitter’s pushbuttons. You should be able to see the 

indicator light (the small window beside the “4” button).

Содержание GM001

Страница 1: ...CONTENTS To the owner Congratulations on receiving your GPI 4 20 mA Out with Display We are pleased to provide you with a product designed to give you maximum reliability and efficiency Our business i...

Страница 2: ...in indoor or outdoor applications where occasional exposure to moisture is common The 4 20 mA Out with Display is available in two ver sions One indicates flowrate in units minute and one in units ho...

Страница 3: ...her liquids occasional low pressure splashing will not harm unit if cable entry points are well sealed subject to 5g shock loading facing the sun directly for long periods of time close to high voltag...

Страница 4: ...m ONLY Local Com HL HF IN R See note EDM Power INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Min V 8 5 Max V 35 Customer Equipment Minimum signal amplitude required for normal oper...

Страница 5: ...Max V 35 Customer Equipment Loop Power Supply Typ 12 24 VDC Loop Input Loop Input NOTE 1 Loop power supply electrically isolated from customer equipment NOTE 2 Actual value of loop current mA is disre...

Страница 6: ...Frequency Generating Input Device INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Customer Equipment Loop Power Supply Typ 12 24 VDC Open Collector Pulse In Pulse In Return Pulse Ou...

Страница 7: ...that the enclosure sealisnotcrimpedortwisted Donotover tightencornerscrews hand tighten only Faceplate can be rotated 90 W I R I N G D I A G R A M 8 4 20 mA or 0 20 mA Output Customer Equipment Witho...

Страница 8: ...4 20 mA and Pulse Output option available only with Reed Switch input W I R I N G D I A G R A M 1 0 4 20 mA or 0 20 mA Output and Pulse Output Customer Equipment With Built in Power Supply Input Reed...

Страница 9: ...button for 3 seconds to reset the Batch Total to zero When fluid is flowing through the meter a small propeller icon is highlighted Flowrate Feature Tousethisfeature pressandreleaseDISPLAYuntilFLOWRA...

Страница 10: ...ser to tweak the factory calibration by a percent that represents application fluid or plumbing differences 1 To calibrate press and hold the CALIBRATE and DISPLAY buttonsforabout3secondsuntilyouseeFL...

Страница 11: ...the transmitter If signal is being received deactivate it by one of the following methods Stop the flow through the line that is being recorded Disconnect the output device on the flowmeter Locking U...

Страница 12: ...presently operating temporarily disable its external power supply Be sure to allow at least 30 seconds to elapse with unit unpowered 2 Press and hold the 4 button While holding watch the indicator li...

Страница 13: ...signal input connections for intermittent open or short circuits F LOOP OUTPUT 1 Output response time setting too long Select a shorter response time setting STABLE BUT especially for fast input sign...

Страница 14: ...14 DIMENSION DRAWINGS...

Страница 15: ...Tapping GM001 GM002 and GM003 2 9 120058 01 Bracket 1 10 904005 13 Screw 6 32 x 1 2 in 4 Item No No Part No Description Req d 11 904002 44 Screw 8 32 x 5 16 in 2 12 125066 20 Cable 20 ft 1 125066 3 C...

Страница 16: ...linearity plus span any input loop current output 0 C to 70 C loop voltage supply 12 VDC to 24 VDC 0 5 of span plus possible resolution uncertainty Max Conversion Resolution Uncertainty Loop current o...

Страница 17: ...gital Pulse Out Open Collector square wave Will switch up to 60 V and up to 200 mA Closed circuit voltage drop typically 0 1 V guaranteed less than 0 5 V at any in range current Configuration 2 Totals...

Страница 18: ...Please contact GPI before returning any part It may be possible to diagnose the trouble and find a solution with a telephone call GPI can also inform you of any special require ments you will need to...

Страница 19: ......

Страница 20: ...rranty shall not apply if A the product has been altered or modified outside the warrantor s duly appointed representative B the product has been subjected to neglect misuse abuse or damage or has bee...

Отзывы: