10.2.5 RISO – L-N
This measurement is used to test the insula-
tion of, for example, extension cords and
multiple outlets (via optional adapters) with-
out circuitry (e.g. glow lamps).
Test voltage is applied between L and N to
this end, and insulation resistance is mea-
sured.
Connection
➭
Insert the test object’s mains plug into
the test socket.
➭
Connect the probe to the red probe
socket.
Procedure
➭
Press the
R
ISO
key.
➭
Select the “
L-N
” menu with the help of
the
scroll keys.
➭
Start the measurement by pressing the
OK
key.
➭
Contact all conductive parts with the
probe.
Note!
There’s no limit value for this mea-
surement, and thus there’s no corre-
sponding LED display.