10.1.5 RPE – Balancing the Probe Cable for the
Protective Conductor Measurement
This menu makes it possible to balance the
probe cable’s resistance value.
The maximum offset value, which is sub-
tracted from subsequent measured values,
must not exceed 0.100
.
Connection
➭
Connect the probe to the red probe
socket.
Procedure
➭
Press the
R
SL
key.
➭
Select the “
Rsl Balancing
” menu with the
help of the
scroll keys.
➭
Start the measurement by pressing the
OK
key.
➭
Contact one of the test socket’s two
protective conductor terminals.