10.2.4 RISO – Probe-Probe (GND)
Insulation Resistance Measurement, PC III
This measurement is intended exclusively for
protection class III DUTs.
Insulation resistance between two points
(e.g. components) can be determined with
this measurement.
The test voltage and the limit value are
adjustable.
This measurement is not performed if there
are no insulated, conductive parts.
Conductive insulated parts must be con-
tacted one after the other.
Measurement is performed with the DUT in
the de-energized state.
Measurement is conducted with a test volt-
age of 250 or 500 V.
Connection
➭
Connect the first probe to the black GND
socket.
➭
Connect the second probe to the red
probe socket.
➭
Connect phase conductors L1, L2 and
L3 to N and the red probe.
➭
Connect PE to the black probe.
Procedure
➭
Press the
R
ISO
key.
➭
Select the “
Probe-Probe
” menu with the
help of the
scroll keys.
➭
Start the measurement by pressing the
OK
key.
➭
Review the preconfigured parameters for
the limit value (0.3/1/2 M
) and for test
voltage (250 V/500 V), select them if
necessary using the
scroll keys and
adjust them to the momentary values
using the
scroll keys.
Heating Elements:
“Yes” results in a reduction
of the permissible limit value of 1 M
to
0.3 M
.
Test Voltage:
In the case of devices with inte-
grated surge arrester, insulation resistance
measurements performed with 500 V may
lead to
“excessively small” measurement results.
Reducing test voltage to 250 V makes it
possible to perform measurement without
activating overvoltage protection.
➭
Contact the short-circuited L and N pins
in the DUT’s mains plug with the first
probe.
➭
Contact all of the DUT’s insulated, con-
ductive parts with the second probe.
➭
If the respective measured value is larger
than the specified limit value,
OK
appears
at the display and the
LED lights up
green.
➭
You can repeat the measurement by
pressing the
scroll key (reset), after
which the LED
goes out briefly until
the next measurement is performed.